Ultimate Healthcare Heritage II Bariatric User Manual

Ultimate Healthcare Heritage II Bariatric User Manual

Bariatric dynamic mattress replacement system

Advertisement

Quick Links

Heritage II Bariatric
Bariatric Dynamic Mattress Replacement System
UPRA4878-2
User Manual
Issue 5: 12/12/2018
UPRA4878-2.UM-5

Advertisement

Table of Contents
loading

Summary of Contents for Ultimate Healthcare Heritage II Bariatric

  • Page 1 Heritage II Bariatric Bariatric Dynamic Mattress Replacement System UPRA4878-2 User Manual Issue 5: 12/12/2018 UPRA4878-2.UM-5...
  • Page 2 Issue: 5 - 12/12/2018 Page 2 of 28...
  • Page 3: Table Of Contents

    Contents Introduction ............................5 Intended Use ............................ 5 2.1 Contraindications ........................5 About the Product ..........................5 Symbols and Statements ......................... 7 Important Safety Information ......................8 Technical Specification ........................9 6.1 Power unit ..........................9 6.2 Mattress ........................... 9 Installation and Set-Up........................
  • Page 4 16. Service and Maintenance ......................21 16.1 General ..........................21 16.2 Fuse replacement ........................21 16.3 Air filter replacement ......................21 17. EMC Information ..........................22 Issue: 5 - 12/12/2018 Page 4 of 28...
  • Page 5: Introduction

    Bariatric provides comfortable care options for larger patients considered to be at High to Very High Risk of pressure damage. At 1200mm wide the Heritage II Bariatric features a full width deep 8” alternating cell design to provide optimum support and pressure care for larger patients. Automatic features optimise pressure area care and patient protection whilst reducing the burden on nursing staff.
  • Page 6 conform closely a patient’s body and redistribute body weight over a wider area; this greater patient immersion delivers optimum constant levels of pressure redistribution. Mattress Replacement 3 x Static Head Cells (TPU) HEAD END Knitted Cell Retention Bands Removable Cover 1 x Standard Cell (TPU) CPR Tag 11 x Standard Cells (TPU)
  • Page 7: Symbols And Statements

    4. Symbols and Statements Note: Indicates tips and advice for the correct use of this product. Caution: Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the product or other property. Warning: Indicates potential danger that requires correct procedures or practises in order to prevent personal injury.
  • Page 8: Important Safety Information

    Use only the cleaning and disinfectant agents recommended in this User Manual. When connecting product after transportation or storage, inspect the power cable visually for any signs of damage. If evident, do not use product and contact Ultimate Healthcare or your local distributor for repair.
  • Page 9: Technical Specification

    6. Technical Specification Power unit Dimensions: 130mm x 360mm x 230mm (D x W X H) Weight: 4.6 kg Alternating cycle time: 10 / 15 / 20 / 25 / 30 mins Output pressure range: 25 to 60mmHg (+/-2) Power supply: AC 230V 50 Hz Current: 0.12 A...
  • Page 10: Installation And Set-Up

    7. Installation and Set-Up Setting up mattress For the comfort and safety of the patient do not put them onto the mattress until you are sure that the mattress is properly secured and the system indicates that it is fully inflated. The mattress is designed to completely replace any existing mattress which may be in use on a bed.
  • Page 11: Cable Management System

    Cable management system Cable management fixings are located on each side of the mattress underneath the flap of the cover. The mains power cable should be secured through the cable management fixing as follows:  Locate each cable management fixing. ...
  • Page 12: Control Panel Operation Guide

    8 Control Panel Operation Guide The Control Panel of the power unit is used to make adjustments to the mattress and also indicates fault conditions/service requirements. These are either visual (indicator lights) or audible. When the incorrect button is pressed a buzzer will sound to indicated that the pump is responsive. Auto detection Auto-Firm &...
  • Page 13: Comfort Level

    Comfort level The Soft and Firm buttons allow carers to adjust pressures within a safe pre-set range to provide patients with enhanced comfort or support whilst maintaining a very good level of protection and therapy. Qualified clinical advice must always be taken before adjusting mattress pressures.
  • Page 14: Function Mode Switch

    Function mode switch 8.5.1 Alternate Alternating mode is the default mode for the system. Within this mode the mattress will operate in an alternating 1-in-2 cell cycle. The alternating cycle will continue at the selected cycle time until another mode is selected. 8.5.2 Static Pressing the Function Mode Selection button until the Static indicator illuminates puts the system into Static mode.
  • Page 15: Backrest Mode

    This will maintain the cells in their present state for approximately 48 hours. It is important to restore the Heritage II Bariatric mattress as quickly as possible by reconnecting the supply tubes to the power unit. Issue: 5 - 12/12/2018...
  • Page 16: Alarms & Fault Findings

    12 Alarms & Fault Findings The Heritage II Bariatric is equipped with audible and visual alarm indicators. These alert the user to the status of the available mains supply and any mattress defect. 12.1 Low pressure alarm Upon detection of low pressure, an audible alarm will be heard and the Low Pressure indicator will illuminate.
  • Page 17 Table 1: Alert/Error Code Reference Table Priority High Warning Indicator Audible Condition of Warning Description Remarks ↓ Code Output Mode Output Not in System Key Tone from ONCE Key Tone Shutdown Functional Button Power ONCE Power-Off System Shutdown Shutdown Failure Operate or All Indicators ALL LED...
  • Page 18: Troubleshooting

     Indicates that system is scheduled for an annual Service indicator remains service. Please contact Ultimate Healthcare or your illuminated local dealer to arrange. If the problem persists, contact Ultimate Healthcare or your local service provider. Issue: 5 - 12/12/2018 Page 18 of 28...
  • Page 19: Cleaning And Decontamination

    14.2 Mattress and cover disinfection The Heritage II Bariatric mattress, mattress cover and air pipe cover can be cleaned using the following simple procedures in accordance with your Local Infection Control Policy: ...
  • Page 20: Power Unit Disinfection

    14.3 Power unit disinfection The power unit can be cleaned by wiping down with a cloth dampened with hot water at 60 containing detergent or with sodium hypochlorite (up to 10,000 parts per million available chlorine). 14.4 Cover laundering The mattress cover and air pipe cover can also be machine washed. Mattress covers and air pipe cover should be completely removed prior to laundering.
  • Page 21 12 months regardless of product usage. The Heritage II Bariatric features a dedicated Annual Service indicator on the control panel which will illuminate and stay ON to alert nursing staff of the need for the system to be professionally serviced.
  • Page 22 17 EMC Information Guidance and manufacturer’s declaration-electromagnetic emissions The GD311-401 is intended for use in the electromagnetic environment specified below. The customer or the user of the GD311-401 should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment-guidance RF emissions...
  • Page 23 Guidance and manufacturer’s declaration-electromagnetic immunity The GD311-401 is intended for use in the electromagnetic environment specified below. The customer or the user of the GD311-401 should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level Electromagnetic test level environment-guidance...
  • Page 24 Guidance and manufacturer’s declaration-electromagnetic immunity The GD311-401 is intended for use in the electromagnetic environment specified below. The customer or the user of the GD311-401 should assure that is used in such and environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance Portable and mobile RF communications...
  • Page 25 Recommended separation distance between portable and mobile RF communications equipment and the GD311-401 The GD311-401 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the GD311-401 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the GD311-401 as recommended below, according to the maximum output power of the communications equipment.
  • Page 26 Issue: 5 - 12/12/2018 Page 26 of 28...
  • Page 27 Issue: 5 - 12/12/2018 Page 27 of 28...
  • Page 28 Ultimate Healthcare Ltd Calmore Industrial Estate, Nutwood Way, Totton Southampton, Hampshire, SO40 3WW Tel: 0333 321 8996 Fax: 023 8066 2388 Email: info@ultimatehealthcare.co.uk www.ultimatehealthcare.co.uk Issue: 5 - 12/12/2018 Page 28 of 28...