Tamora Plus II Digital
Dynamic Mattress Replacement System
UPRA367803D-2
_______________________________________________________
User Manual
Issue 9 – 15/09/2022
UPRA367803D-2.UM-8

Advertisement

Table of Contents
loading

Summary of Contents for Ultimate Healthcare UPRA367803D-2

  • Page 1 Tamora Plus II Digital Dynamic Mattress Replacement System UPRA367803D-2 _______________________________________________________ User Manual Issue 9 – 15/09/2022 UPRA367803D-2.UM-8...
  • Page 2 Page 2 of 28 Issue 9 – 15/09/2022...
  • Page 3: Table Of Contents

    Contents Introduction ............................5 Intended Use ............................ 5 2.1 Contraindications ........................5 About the Product ..........................5 Symbols and Statements ......................... 7 Important Safety Information ......................8 Technical Specification ........................9 6.1 Power unit ..........................9 6.2 Mattress ........................... 9 Installation and Set-Up........................
  • Page 4 15.2 Fuse replacement ........................21 15.3 Air filter replacement ......................21 16 EMC Information ..........................22 Page 4 of 28 Issue 9 – 15/09/2022...
  • Page 5: Introduction

    User Manual. Note: Ultimate Healthcare reserves the right to modify the information in this User Manual at any time. The information in this User Manual may vary slightly with respect to the basic design of the product.
  • Page 6 Mattress Replacement HEAD END 3 x Static Head Cells (TPU) Knitted Cell Retention Bands Removable Cover CPR Tag 2” Foam Base Compartment 13 x Standard Cells (TPU) FOOT END Cable Management Fixings Air Pipe Connector Power Unit Carry Handle Control Panel Swing Out Hooking Hooks Air Pipe...
  • Page 7: Symbols And Statements

    4. Symbols and Statements Note: Indicates tips and advice for the correct use of this product. Caution: Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the product or other property. Warning: Indicates potential danger that requires correct procedures or practises in order to prevent personal injury.
  • Page 8: Important Safety Information

    If this product is used for any activity other than detailed within this User Manual then personal risk to the end user or patient may occur. Ultimate Healthcare shall not be held liable/responsible for such an event. Do not smoke, or allow the patient to smoke when using this product. Keep all possible ignition sources...
  • Page 9: Technical Specification

    6. Technical Specification Power unit Dimensions: 130mm x 320mm x 230mm (D x W X H) Weight: 3.5 kg Alternating cycle time: 10 / 15 / 20 / 25 / 30 mins Output pressure range: 25 to 60mmHg (+/-2) Power supply: AC 230V 50 Hz Current: 0.12 A...
  • Page 10: Installation And Set-Up

    7. Installation and Set-Up Setting up mattress For the comfort and safety of the patient do not put them onto the mattress until you are sure that the mattress is properly secured and the system indicates that it is fully inflated. The mattress is designed to completely replace any existing mattress which may be in use on a bed.
  • Page 11: Cable Management System

    Cable management system Cable management fixings are located on each side of the mattress underneath the flap of the cover. The mains power cable should be secured through the cable management fixing as follows:  Locate each cable management fixing. ...
  • Page 12: Control Panel Operation Guide

    Table 1: Weight and Suggested Comfort Level Reference Table Pump output Patient Weight (kg) Comfort Control Pressure (Auto-Detection) (mmHg) ● ● ● ● ● ● ● ● < 60 ● ● ● ● ● ● ● ● 40 – 80 ●...
  • Page 13: Operate / Standby

    Operate / standby Press the Operate / Standby button to turn the power unit ON. Press again to turn OFF/Standby the power unit.  Note: The power switch on the side of the power unit must be turned ON. Comfort level The Soft and Firm buttons allow carers to adjust pressures within a safe pre-set range to provide patients with enhanced comfort or support whilst maintaining a very good level of protection and therapy.
  • Page 14: Control Panel Lockout

    Control panel lockout If the control panel is not used for a period of 30 seconds it will lock out and inhibit the use of the functions. Additionally, if you wish to lock out the control panel press the Lock Out button and all functions will be locked. This is to prevent the system being altered accidently.
  • Page 15: Alternate Cycle Time Selection

    Alternate cycle time selection The alternating cycle times can be selected to provide an individualised care program for each patient. Within Alternating mode the alternating cycle time can be selected by pressing the Cycle button. Selections can be made from 10-30 minutes at 5 minute intervals. The cycle time will be displayed in the LCD screen.
  • Page 16: Alarms & Fault Findings

    Should your system develop a fault condition whilst in use, an audible alarm will be heard and the Service indicator will illuminate. The audible alarm may be cancelled by pressing the Alarm Mute button. Please refer to Table 2 for error codes and call Ultimate Healthcare or your local distributor. 11.4 Annual Service Indicator A dedicated Annual Service indicator on the control panel will illuminate and stay ON to alert nursing staff of the need for the system to be professionally serviced.
  • Page 17 Table 2: Alert/Error Code Reference Table Priority High Warning Indicator Audible Condition of Warning Description Remarks ↓ Code Output Mode Output Not in System Key Tone from ONCE Key Tone Shutdown Functional Button Power ONCE Power-Off System Shutdown Shutdown Failure Operate or All Indicators ALL LED...
  • Page 18: Troubleshooting

     Indicates that system is scheduled for an annual Service indicator remains service. Please contact Ultimate Healthcare or your illuminated local dealer to arrange. If the problem persists, contact Ultimate Healthcare or your local service provider. Page 18 of 28 Issue 9 – 15/09/2022...
  • Page 19: Cleaning And Decontamination

    Healthcare cannot be held liable for.  Caution: The appropriate qualified staff must be consulted when specifying a suitable cleaning fluid. Ultimate Healthcare shall not be liable for any damages caused by the use of inappropriate detergents or disinfectants. 13.2 Mattress and cover disinfection...
  • Page 20: Power Unit Disinfection

     Caution: Frequent or prolonged exposure to higher concentration disinfectant solutions may prematurely age the fabric cover of mattresses. Surfaces must be protected during use and rinsed and thoroughly dried after application of a disinfectant. 13.3 Power unit disinfection The power unit can be cleaned by wiping down with a cloth dampened with hot water at 60 containing detergent or with sodium hypochlorite (up to 10,000 parts per million available chlorine).
  • Page 21: Service And Maintenance

    15 Service and Maintenance It is recommended that this product be part of a routine preventative maintenance schedule with a planned service every 12 months regardless of product usage. The Tamora Plus II Digital features a dedicated Annual Service indicator on the control panel which will illuminate and stay ON to alert nursing staff of the need for the system to be professionally serviced.
  • Page 22 EMC Information Manufacturer’s declaration-electromagnetic emissions The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the device(s) should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment-guidance...
  • Page 23 Manufacturer’s declaration-electromagnetic immunity The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the device(s) should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level Electromagnetic environment-...
  • Page 24 Manufacturer’s declaration-electromagnetic immunity The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the device(s) should assure that it is used in such and environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance...
  • Page 25 Recommended separation distance between portable and mobile RF communications equipment and the device(s) The device(s) is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF disturbances are controlled. The customer or the user of the device(s) can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device(s) as recommended below, according to the maximum output power of the communications equipment.
  • Page 26 Manufacturer’s declaration-electromagnetic immunity Test specifications for to RF wireless communications equipment ENCLOSURE PORT IMMUNITY The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the device(s) should assure that it is used in such an environment. Compliance LEVEL Test...
  • Page 27 Notes: Page 27 of 28 Issue 9 – 15/09/2022...
  • Page 28 Ultimate Healthcare Ltd Calmore Industrial Estate, Nutwood Way, Totton Southampton, Hampshire, SO40 3WW Tel: 0333 321 8996 Fax: 023 8066 2388 Email: info@ultimatehealthcare.co.uk www.ultimatehealthcare.co.uk Page 28 of 28 Issue 9 – 15/09/2022...