Trouble Shooting

16
The table below has been created to assist with resolving faults you may come across; if the instructions here
do not resolve the issue or the issue you are experiencing is not listed please contact your supplier or Ultimate
Healthcare using the details found on the back cover of this manual.
 No adjustment of the bed
can be made when the
power is ON
 The function shows orange
on the handset.
 The handset doesn't
function –
 When a button is pressed
on the handset, the
corresponding actuator
does not respond
 An unexpected operation
occurs when a button is
pressed on the handset
Issue 3: 20/04/20
Problem
Possible Cause
Handset function is locked.
The function is locked out.
Mains power not
connected
Handset not connected to
the controller
Connectors not plugged
into the correct sockets
Faulty actuator
Connectors not plugged
into the correct sockets
Resolution
Release the handset function
using the key
Unlock the function if
required and try again.
Connect to the mains power
Connect the handset to the
controller; following the
instructions in this manual.
Check the connections, are
as per the instructions in this
user guide
Contact your supplier or
Ultimate Healthcare using
the details on the back cover
of this manual
Check the connections, as
per the instructions in this
user guide
Contact your supplier or
Ultimate Healthcare using
the details on the back cover
of this manual
Page 23 of 32