SECTION 8. TROUBLESHOOTING

DO NOT attempt to service the Sonic Ruptor 400 in a manner other than those discussed in
this manual. For any issue that is unsuccessfully corrected using this guide, please contact your
authorized dealer or call Omni International at 1-800-776-4431.

8.1 INSTALLATION

Error Message
Unit will not
power on

8.2 OPERATION

Problem
Power output is
less than 20%
Burning smell from
transducer
Possible Causes
Control Unit is not plugged into
a functional electrical outlet.
The transducer is not properly
connected to the control unit.
Fuse needs to be replace.
Possible Cause
No Power or
incorrect voltage
Worn washer
The probe is not properly
secured
The transducer is overheating
Action(s)
1. Make sure the unit is plugged in properly.
2. Check that the voltage of the unit match-
es that being delivered by the main power
supply.
1. Check connection and secure properly.
1. Check and replace fuses if blown.
Action(s)
1. Check main voltage.
2. Check that the voltage of the unit
matches that being delivered by the main
power supply.
3. Make sure the unit is plugged in properly.
1. Turn off the unit.
2. Use the wrenches to loosen the
processing tip from the transducer.
Be sure to twist the tip, not the
transducer.
3. Replace washer
1. Turn off the unit. Use the wrenches to
tighten the processing tip on the
transducer.
1. Turn off the control unit.
2. Wait at least 30 minutes before turning
the unit back on.
3. Contact Technical Support if the problem
persists.
1. Replace the washer.