5.1 OPERATION

An understanding of the operation of each component of the Omni Sonic Ruptor 400 is
important before applying the instrument to an application.
Transducer
The Transducer assembly may be hand held or mounted to a support stand (sold separately).
To test the Transducer, fill a small beaker half full of water. Turn the device on. The LED on
the front panel will light up. Immerse the Tip in the water approximately a centimeter. Turn
the Power Control Knob to a high range. Push the Start button. Intense cavitation will occur
in the water. No warm-up is required.
CAUTION: If using a stand clamp, tighten the clamp to the white portion of the Transducer,
NOT to the Processing Tip.
WARNING: DO NOT touch the titanium Processing Tip while power is being delivered to
the tip.
WARNING: The Transducer and Tip will become hot after long periods of operation.
Ready To Use
After the initial testing of the instrument, push the "Reset" button.

Set the desired Power Level, Pulser, and Timer.

Immerse the Tip in the appropriate solution. Tip depth should be

approximately equal to 1.5 times the diameter of the tip.
Push the "Start" button to deliver the ultrasonic power to the

convertor.
If the Timer is used, the processing will automatically stop when

the Timer is through.
Push the "Reset" button to terminate the processing at any time.


SECTION 5. OPERATION