Model 2461 Calibration and Adjustment Manual

:CALibration:DIGitize:ADJust:SENSe

This command adjusts the Model 2461 digitizer function.
Type
Query only
Usage
:CALibration:DIGitize:ADJust:SENSe
Details
This command must be sent after all other :CALibration:ADJust commands for all ranges have
been completed successfully.
The Model 2461 does not source any output and there is no need to connect external equipment
during this adjustment step.
This command initiates an internal operation to adjust digitizer circuits to match the readings of
integration analog-to-digital circuits.
This command requires about 10 seconds to complete.
Example
-- Perform all other adjustment steps
first.
:CAL:DIG:ADJ:SENS; *OPC?
Also see
:CALibration:ADJust:SENSe
:CALibration:ADJust:SOURce
2461-905-01 Rev. A / February 2016
Affected by
Not applicable
(on page 5-3)
(on page 5-5)
Where saved
Not applicable
The *opc? query returns "1" when the
adjustment command is finished. This command
may require a longer timeout value than most
commands.
Section 5: Command reference
Default value
Not applicable
5-7