Section 4: Adjustment

Single-range adjustment

Normally, the complete adjustment procedure should be performed to ensure that the entire
instrument is properly adjusted. In some instances, however, you may want to adjust only certain
ranges.
To adjust only some of the ranges, complete the entire unlock, adjust, save, and lock procedure for
the range or ranges to be adjusted. Note that you must complete all four source adjustment steps and
all three sense adjustment steps for each range to be adjusted. Also, be sure to set calibration
verification and adjustment dates and save calibration constants after adjusting the ranges.
4-18
Model 2461 Calibration and Adjustment Manual
2461-905-01 Rev. A / February 2016