Section 5: Command reference
:CALibration:ADJust:SOURCe:DATA?
This command queries the calibration constants for the active source range.
Type
Query only
Usage
CALibration:ADJust:SOURce:DATA?
Details
This command requests the calibration constants for the active range of the source function.
Constants returned are the positive full-scale value, the positive zero value, the negative full-scale
value, and negative zero value.
The returned constants are in comma-delimited ASCII floating-point format.
Example
:SENS:FUNC CURR
:SENS:VOLT:RANG 1
:CAL:ADJ:SOUR:DATA?
Also see
:CALibration:ADJust:SOURce
5-6
Affected by
Not applicable
(on page 5-5)
Model 2461 Calibration and Adjustment Manual
Where saved
Not applicable
Select the current source function.
Select the 1 A range.
Request the calibration constants.
Example output:
+3.018486E+05,+1.297160E+03,
-3.018662E+05,+1.248219E+03
2461-905-01 Rev. A / February 2016
Default value
Not applicable