Model 2461 Calibration and Adjustment Manual
Current calibration adjustment data table
Source range
1 μA
10 μA
100 μA
1 mA
10 mA
100 mA
1 A
* DMM reading used in corresponding adjustment step. See the adjustment procedure in
Remote adjustment procedure
2461-905-01 Rev. A / February 2016
Source current
DMM current reading*
+1.0000 μA
+0.0000 μA
−1.0000 μA
−0.0000 μA
+10.000 μA
+00.000 μA
−10.000 μA
−00.000 μA
+100.00 μA
+000.00 μA
−100.00 μA
−000.00 μA
+1.0000 mA
+0.0000 mA
−1.0000 mA
−0.0000 mA
+10.000 mA
+00.000 mA
−10.000 mA
−00.000 mA
+100.00 mA
+00.000 mA
−10.000 mA
−00.000 mA
+100.00 mA
+000.00 mA
−100.00 mA
−000.00 mA
(on page 4-7).
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ μA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ mA
____________________________ A
____________________________ A
____________________________ A
____________________________ A
Section 4: Adjustment
4-5
Need help?
Do you have a question about the KEITHLEY SourceMeter 2400 Series and is the answer not in the manual?