10. PROGRAMMING A PRESET BUTTON WITH A SINGLE FILE
1. Disable the Preset Memory Lock.
For instructions, read about the System
Setup Menu beginning on page 6.
2. Press and hold the desired Preset Button for
1 second (until you hear a double beep).
3. The display will show:
ACCESS FILE LIBRARY?
1=yes, 2=no, 3=esc
Press Button #1 to access the File Library.
4. The display will show:
PATHFILE ALL
1=next, 2=ok, 3=esc
Press Button #1 repeatedly until the desired
file series is shown on the display. See
Chart 3 on page 11 for a list of available file
series.
Example 2:
To go to the Protaper
press preset button #1 three times.
The display will then show:
PROTAPER UNIVERSAL
1=next, 2=ok, 3=esc
5. Press Button #2 to select the displayed file
series.
6. The display will show:
LOAD ENTIRE SERIES?
1=yes, 2=no, 3=esc
Press Button #2. Do not load entire series.
7. Press Button #1 repeatedly until the desired
file is shown on the display.
See Chart 3 on page 11 for a list of available
files in each file series.
12
®
Universal series,
Example 2:
To go to the F2 and F3 files in the
Protaper series, press Button #1 twice.
The display will show the following:
PROTAPER F2 F3 F4 F5
1=next, 2=ok, 3=esc
8. Press Button #2 to select the displayed files.
9. The display will show:
USE DEFAULT HP?
1=yes, 2=no, 3=esc
NOTE: The default handpiece is the
handpiece that was shipped with the Endo
DTC
®
, specified in Quick Start Operation
Section 2.
a. If you are using the default handpiece
with this file series, press Button #1 and
go on to Step 10.
b. If you are using a handpiece other than
the default handpiece, press Button #2
and choose the correct handpiece. For
instructions, refer to i and ii on page 10.
10. The Preset Button is now set with the
selected single file settings.
For Example 2 above:
The display will now show:
SINGLE FILE PRESET
PROTAPER F2 F3 F4 F5
11. When you have completed the program-
ming of the desired preset buttons, re-
enable the Preset Memory Lock to avoid
inadvertently entering the File Library dur-
ing a procedure. For instructions, read
about the System Setup Menu beginning
on page 6.