CHANGING THE FUSE OR VOLTAGE -
WARNING
Turn the power off and unplug the
unit before following the steps below.
1.
Use a small flathead screwdriver to open the
Fuse/Voltage Selector Panel on the back of
the unit.
2.
Remove the red Fuse holder from the unit.
3a. Changing the fuse:
Replace the blown fuses on either side
of the Fuse holder.
Replacement Fuses:
115V: 1.0A/250V slo-blo fuse
230V: 0.5A/250V slo-blo fuse
3b. Changing the voltage:
The proper fuses for the country of use are
sent in the unit. When changing voltage,
replace the fuses with fuses that are rated
for the new voltage.
Turn the Fuse holder until the white lettering
with the proper voltage is right side up.
4.
Replace the Fuse holder and close the
Fuse/Voltage Selector Panel.
5.
Select a power cord or grounded plug
adapter specific to the country of use.
CAUTION
Do not attempt to operate the Endo DTC
while configured in the 115V mode with a
230V power source. This could damage
the electrical components and will void
the warranty.
CHANGING THE VOLTAGE
Follow the corresponding steps.
®
1.
2.
3b.
4.