Chapter 3
Meter Setup Menus
|
24
Orion Star A220 Series Portable Meter User Manual
|
Mode and Settings Menu for COND Channel
Use the Mode and Settings menu to customize settings for the conductivity channel of Orion
Star A222 conductivity meters. Default meter settings are bold.
Parameter
Settings
· Conductivity
· Salinity
Measure Mode
· TDS
· Resistivity
· Auto
Read Type
· Continuous
Cell K
Cell K
TDS Factor
TDS Factor
(TDS only)
· 15°C
Ref. Temp.
· 20°C
· 25°C
· Off
Temp. Comp.
· Linear
· nLFn
Temp. Coeff.
Temp. Coefficient
Cal Due Alarm
· Cal Due Alarm
(Conductivity
only)
· Standard
Cell Type
· USP
Cell K Setting
Orion Star A222 conductivity meters are capable of automatically recognizing Thermo Scientific
Orion 100 μS conductivity standard, 1413 μS conductivity standard and 12.9 mS conductivity
standard when the nominal cell constant (K) value of the conductivity sensor is entered in the
setup menu.
Catalog Number
Description
011008
Orion 100 μS conductivity standard, 5 x 60 mL
011007
Orion 1413 μS conductivity standard, 5 x 60 mL
01100710
Orion 1413 μS conductivity standard, 10 individual use pouches
011006
Orion 12.9 mS conductivity standard, 5 x 60 mL
01100610
Orion 12.9 mS conductivity standard, 10 individual use pouches
Details
Set the displayed measurement mode, the selected mode
determines which additional settings are shown and the
type of calibration the meter will perform
Set the read type to determine how the meter takes a
measurement and when the measurement is sent to the
data log and/or external device
Enter the nominal cell constant (K) value of the
0.4750
conductivity sensor for automatic conductivity calibration
Enter the TDS factor value for total dissolved solids
0.49
measurements
Set the reference temperature for temperature
compensated measurements, readings are adjusted to
the selected reference temperature when the temperature
compensation setting is active
Set the temperature compensation type as linear, nLFn
(non linear natural water) or off
Enter the temperature coefficient used with the linear
2.10
temperature compensation setting
Set the calibration due alarm off or on, when the alarm is
on, the fixed calibration time interval is 12 hours and the
On
alarm is triggered after 12 hours has elapsed without a
calibration being performed
Set the type of conductivity senor as standard (most
sensors) or USP (2-cell sensors in ultra pure water)
Thermo Scientific