Marker Out Settings

The Marker Out settings parameters are located in the Device Setting page. The MARKER OUT BNC
connector is shared by each channel of the instrument but it's possible to link the marker signal coming
out of it to a predetermined one.
Marker Out Setting
Voltage Level [V]
State
Skew [s]
Link To
Description
it sets the Marker high level Voltage. The Low level is fixed to 0V.
The Voltage Level range is 1V to 2.5V on 50Ohm load (twice on open
load).
It enables or disables the Marker Out. When the Marker Out is disabled it
is forced to 0V.
It sets the delay between the marker and the analog channels.
The skew could assume negative values and the absolute range depends
on the Run Mode: in Gated (Burst) it is 3us; for all others modes it is 14s.
The Marker Out frequency is related to the analog channel frequency.
Since there is a unique Marker Out signal while there are two/four analog
channels it is possible, by means of this parameter, to define the analog
channel that is related to the Marker Out signal.
Operator's Manual Simple AFG Application
57