Table 2.4 - Temperature Controller - Initial Settings; Table 2.5 - Temperature Controller - Secondary Menu - Teledyne 6200A Manual

Trs analyzer
Table of Contents

Advertisement

PRESS
DISPLAY
SEL
P PROP BAND
SEL
i INTEGRAL
SEL
d DERIVATIVE TIME
SEL
AL LOW ALARM
SETPOINT
SEL
AH HIHG ALARM
SETPOINT
SEL
7C CYCLE TIME
SEL
HYS HYSTERESIS
SEL
A7 AUTOTUNE
SEL
LOC LOCK
Table 2.4 shows a Secondary Menu of parameters that set more basic parameters of the controller, these
include the thermocouple type, the temperature units etc.
PRESS
DISPLAY
SEL
HOLD TILL p-n1
SEL
p-n2
SEL
p-dF DIGITAL FILTER
SEL
P-SL LOWER LIMIT
SEL
P-SU UPPER LIMIT
SEL
P-AL ALARM TYPE2
SEL
P-AH ALARM TYPE 1
SEL
P-An HYTERESIS
P-dP DECIMAL LOCATION
SEL
PVOF PROCESS OFFSET
SEL
SEL
SVOF SET POINT OFFSET
SEL
P-F
SEL
FUZZY FUZZY LOGIC
Table 2.4 – Temperature Controller - Initial Settings
Table 2.5 – Temperature Controller - Secondary Menu
TAI - Thermal Converter Operator Manual - Page 16
INITIAL VALUE
UP/DOWN SET TO "11"
SET TO "10"
SET TO "7.7"
SET TO "50" (C BELOW FINAL SETPOINT)
SET TO "50" (C ABOVE FINAL SETPOINT)
SET TO "2"
SET TO "3"
SET TO "0" (OFF)
"0" (OPEN) "1" (LOCKED)
"2" (SV ONLY OPEN)
PARAMETER VALUE
SET TO "0"
SET TO "3" (TYPE K THERMOCOUPLE)
SET TO "5"
SET TO "32" (32C)
SET TO "1050" (1050C)
SET TO "900"
SET TO "1050"
SET TO "3"
SET TO "0"
LEAVE AT "0"
LEAVE AT "0"
SET TO "ºC" (CENTIGRADE)
SET TO "ON"