Teledyne 328 Instruction Manual

oxygen analyzers

Advertisement

P/N M33776
6/29/95
i
Teledyne Analytical Instruments

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne 328

  • Page 1 P/N M33776 6/29/95 Teledyne Analytical Instruments...
  • Page 2 Specific Application Information for the in- strument with which this manual is supplied is on page iv just prior to the Table of Con- tents. Teledyne Analytical Instruments...
  • Page 3 Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
  • Page 4 Instrument Model Number: Instrument Serial Number: Micro-Fuel Cell Class: Accuracy: Response and Recovery: Operating Temperature Range: Range of Analysis: Output Signal Voltage: Output Signal Current: Alarm Setpoint #1: Alarm Setpoint #2: Special Features (may be covered by addendum.): Teledyne Analytical Instruments...
  • Page 5: Table Of Contents

    Install Sensor and Test Alternate Sensor ......3.12 Range Set-Up ..............3.14 Set-up Alarms ............... 3.15 Logger Set-Up ..............3.17 Set-Up Clock Functions ............3.19 Changing Passwords For Remote Monitoring and Control ..........3.21 System Statistics ..............3.23 Teledyne Analytical Instruments...
  • Page 6 4.5 Installation Checklist ..............4.10 5 Maintenance & Troubleshooting 5.1 Routine Maintenance ..............5.1 5.2 Troubleshooting ................5.2 Appendices A.1 Specifications Sheet ..............A.1 A.2 Spare Parts List ................A.3 A.3 Material Safety Data Sheet ............A.4 Teledyne Analytical Instruments...
  • Page 7: Sample System

    Introduction The Teledyne Analytical Instruments (TAI) Model 326 series (Models 326, 327, and 328) analyzers use a unique Micro-Fuel Cell to measure the concentration of oxygen in a gas stream. The analyzer can be housed in a compact equipment case, a 19-inch rack mount unit, an explosion-proof bulkhead-mounted enclosure, or separately from the panel-mounted control unit.
  • Page 8: Model Designations

    See Specific Application Information for your instrument in front of the Table of Contents. 1.1 Model Designation The basic model number of each of the analyzers (326, 327, and 328) has a suffix that describes the options included. Suffixes for the standard options are listed below.
  • Page 9: Models 327R Description

    Introduction Introduction Model 326/327/328 Model 326/327/328 Introduction Introduction Introduction Model 326/327/328 Model 326/327/328 Model 326/327/328 diagram is then included with the drawings in the instruction manual to illustrate the sample flow path. 1.3 Models 327R Description Model 327R is a semi-explosion-proof version of the 326R series. The specific explosion-proof design is dependent on area classification and the specific installation.
  • Page 10: Models 328R Description

    Before supplying power to the analyzer, be sure to check the wiring Before supplying power to the analyzer, be sure to check the wiring Before supplying power to the analyzer, be sure to check the wiring Before supplying power to the analyzer, be sure to check the wiring Teledyne Analytical Instruments...
  • Page 11 Introduction Introduction Model 326/327/328 Model 326/327/328 Introduction Introduction Introduction Model 326/327/328 Model 326/327/328 Model 326/327/328 against the interconnection diagram. against the interconnection diagram. against the interconnection diagram. against the interconnection diagram. against the interconnection diagram. 1.7 Standard Features The following features apply to the standard 326 series analyzers. Since...
  • Page 12 Having an available spare part is usually more economical the services of highly trained service personnel. The printed circuit assemblies are preset and adjusted to be instantly interchanged. 1.8 Optional Features The following optional features are only available at the time the order is placed. Teledyne Analytical Instruments...
  • Page 13 Introduction Introduction Model 326/327/328 Model 326/327/328 Introduction Introduction Introduction Model 326/327/328 Model 326/327/328 Model 326/327/328 1.5.1 Integral Alarm Circuitry 1.5.1 Integral Alarm Circuitry 1.5.1 Integral Alarm Circuitry 1.5.1 Integral Alarm Circuitry 1.5.1 Integral Alarm Circuitry A number -1 or -2 appended to the model number indicates one or two alarm and/or control relay circuits with setpoints that are adjustable over the selected analysis range.
  • Page 14 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxyg cent Oxygen Anal en Anal en Analyz en Anal yz yz yz yzer er er er er Section 1 Section 1 Section 1 cent Oxyg en Anal Section 1 Section 1 Teledyne Analytical Instruments...
  • Page 15 Operational Theory Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 Model 326/327/328 Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 Operational Theory Operational Theory Operational Theory Operational Theory Operational Theory 2.1 Method of Analysis 2.1 Method of Analysis 2.1 Method of Analysis 2.1 Method of Analysis...
  • Page 16 With the incorporation of the FET in the cell circuit, the cell looks into essentially a short circuit at all times and the instrument responds immediately when power is restored to the circuit. Teledyne Analytical Instruments...
  • Page 17 Operational Theory Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 Model 326/327/328 Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 2.3 Cell 2.3 Cell 2.3 Cell T T T T T ypes ypes ypes ypes 2.3 Cell 2.3 Cell ypes TBE selects the class of cell used in the analyzer to give the best pos- sible performance for the customer’s sample conditions.
  • Page 18 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxyg en Anal en Anal yz yz yz yzer er er er er Section 2 Section 2 cent Oxyg cent Oxygen Anal en Anal en Analyz Section 2 Section 2 Section 2 Teledyne Analytical Instruments...
  • Page 19 Operational Theory Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 Model 326/327/328 Operational Theory Operational Theory Model 326/327/328 Model 326/327/328 2.3.7 CO 2.3.7 CO Effect on C-3 and B-1 Cells Effect on C-3 and B-1 Cells 2.3.7 CO 2.3.7 CO 2.3.7 CO...
  • Page 20 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxyg en Anal en Anal yz yz yz yzer er er er er Section 2 Section 2 cent Oxyg cent Oxygen Anal en Anal en Analyz Section 2 Section 2 Section 2 Teledyne Analytical Instruments...
  • Page 21: Installation

    Installation Installation Installation Installation Installation Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Installation Installation Installation Installation Installation Note: Note: If you have ordered special options and have obtained a copy of If you have ordered special options and have obtained a copy of...
  • Page 22: Electrical Connections

    Then, check the wiring with a voltmeter to be sure that the H and N wires are not transposed at the barrier strip. (The neutral should read zero voltage when measured to ground.) The Teledyne Analytical Instruments...
  • Page 23 Installation Installation Installation Installation Installation Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 ground wire should be securely attached to a cold water pipe so that per- sonnel are afforded maximum short circuit protection. As with all other barrier strip connections, the wiring should be termi- nated in an appropriate lug.
  • Page 24 The NO and NC contacts share the C (for common) contact. Any load connected across the NO and C contacts will always see an open circuit (switched off) when the relay is de-energized. Any load connected Teledyne Analytical Instruments...
  • Page 25: Gas Line Connections

    Installation Installation Installation Installation Installation Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 across the NC and C contacts will always see a closed circuit (switched on) when the relay is de-energized. Each set of relay switch contacts (NC-C-NO) is assigned its own barrier strip terminal inside the control unit.
  • Page 26 ¼-inch tubing is recommended. 2. The vent line must terminate in an area that experiences no more than normal barometric pressure changes. 3. The vent line must be installed so that water and/or dirt cannot accumulate in it. Teledyne Analytical Instruments...
  • Page 27 Installation Installation Installation Installation Installation Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 3.3.4 Sample Conditioning Accessories 3.3.4 Sample Conditioning Accessories 3.3.4 Sample Conditioning Accessories 3.3.4 Sample Conditioning Accessories 3.3.4 Sample Conditioning Accessories The sample must be free of entrained solids and water. A high humidity sample gas, however, is beneficial, since it will prevent water loss from the cell’s electrolyte.
  • Page 28 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxyg cent Oxygen Anal en Anal en Anal en Analyz yz yz yz yzer er er er er Section 3 Section 3 Section 3 cent Oxyg en Anal Section 3 Section 3 Teledyne Analytical Instruments...
  • Page 29 Operations Operations Operations Model 326/327/328 Model 326/327/328 Model 326/327/328 Operations Operations Model 326/327/328 Model 326/327/328 Operations Operations Operations Operations Operations 4.1 Preliminary 4.1 Preliminary 4.1 Preliminary 4.1 Preliminary 4.1 Preliminary Before supplying power to the instrument, TBE suggests that the electri- cal wiring installation and fuses be checked against the interconnection diagram—particularly if the installation has been made by personnel...
  • Page 30 Try to limit the amount of time the Micro-Fuel Cell is exposed to air. Try to limit the amount of time the Micro-Fuel Cell is exposed to air. Try to limit the amount of time the Micro-Fuel Cell is exposed to air. Teledyne Analytical Instruments...
  • Page 31 Operations Operations Operations Model 326/327/328 Model 326/327/328 Model 326/327/328 Operations Operations Model 326/327/328 Model 326/327/328 4.5 Initial Calibration and Equilibration 4.5 Initial Calibration and Equilibration 4.5 Initial Calibration and Equilibration 4.5 Initial Calibration and Equilibration 4.5 Initial Calibration and Equilibration Before stable operation can be guaranteed, the Micro-Fuel Cell will require a period of time to adjust itself to its new environment.
  • Page 32 4.7 Calibration of Alarm Equipped Instruments 4.7 Calibration of Alarm Equipped Instruments 4.7 Calibration of Alarm Equipped Instruments 4.7 Calibration of Alarm Equipped Instruments 4.7 Calibration of Alarm Equipped Instruments Teledyne Analytical Instruments...
  • Page 33 Operations Operations Operations Model 326/327/328 Model 326/327/328 Model 326/327/328 Operations Operations Model 326/327/328 Model 326/327/328 To prevent generating ambiguous alarms during calibration, simply unlock and rotate the set point dial (or dials) until the limits of travel in the NORMAL (rather than ALARM) direction have been attained. After sample flow has been re-initiated, and the proper scale of analysis selected, the alarm point can be precisely reset by simply re-dialing the set point.
  • Page 34 Sample Flow Check the Sample flow daily to make sure that particulate matter is not blocking sample flow. 4.10.2 Calibration 4.10.2 Calibration 4.10.2 Calibration 4.10.2 Calibration 4.10.2 Calibration Span calibration should be checked every two to four weeks. Teledyne Analytical Instruments...
  • Page 35: Routine Maintenance

    Maintenance Maintenance Maintenance Maintenance Maintenance Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Maintenance & Maintenance & Maintenance & T T T T T r r r r r oub leshooting leshooting Maintenance & Maintenance & oubleshooting leshooting leshooting 5.1 Maintenance...
  • Page 36: Troubleshooting

    Check Micro-Fuel Cell. (1) Correct cell type for instrument (2) Orientation in socket correct. (3) Shorting clip removed. (4) Diaphragm undamaged. b. Check printed circuit boards. (1) Installed in proper slots. (2) Inserted fully and making good electrical contact. Teledyne Analytical Instruments...
  • Page 37 Maintenance Maintenance Maintenance Maintenance Maintenance Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 c. Check calibration gasses: zero, span, and sample are appropriate types or formulations for the instrument and cell type. d. Instrument has had time to stabilize.
  • Page 38 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxyg cent Oxygen Anal en Anal en Analyz en Anal yz yz yz yzer er er er er Section 5 Section 5 Section 5 cent Oxyg en Anal Section 5 Section 5 Teledyne Analytical Instruments...
  • Page 39 Appendix Appendix Appendix Appendix Appendix Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Appendix Appendix Appendix Appendix Appendix A.1 Cell A.1 Cell A.1 Cell A.1 Cell A.1 Cell W W W W W arranty Inf arranty Inf arranty Inf...
  • Page 40: Spare Parts List

    (if available) and the model and serial number of the system for which the parts are intended. for which the parts are intended. for which the parts are intended. for which the parts are intended. for which the parts are intended. Teledyne Analytical Instruments...
  • Page 41 Appendix Appendix Appendix Appendix Appendix Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 Model 326/327/328 SEND ORDERS TO : Teledyne Brown Engineering Analytical Instruments I6830 Chestnut Street City of Industry, California 91749-1580 Telephone (818) 961-9221 TWX: (910) 584-1887 TDYANYL COID Or your local representative.
  • Page 42 P P P P P er er er er ercent Oxyg cent Oxyg cent Oxygen Anal cent Oxyg en Anal en Anal en Analyz yz yz yz yzer er er er er Appendix Appendix Appendix cent Oxyg en Anal Appendix Appendix Teledyne Analytical Instruments...
  • Page 44 Date: 1/6/04 Time : 01:03:45 PM Title: G:\MASTER\34000\D3420419.DWG...