Channel Setting
Initial Delay
Load [Ohm]
Attenuator
Noise Level [V]
Internal Noise
Polarity
High Limit
Low Limit
Description
Set the Initial Delay of the selected channel. The delay range is 0s to 14 s.
The instrument applies the appropriate scaling to the output waveform to
get the right amplitude on the defined Load expressed in Ohm.
The impedance range is 1 Ohm to 1MOhm.
Use this control to change the full-scale range of the channel. On the
instrument are available two full scale ranges that are selectable by
means of an internal 10X attenuator. You can set the attenuator ON to
improve the output signal quality with low level signals.
Note: the attenuator is turned off with Low Output Impedance.
Use this setting to add noise to the output signal. The voltage shown in
the textbox defines the peak voltage of the noise level. The range is 0V to
6Vpk. The Noise is hardware generated using a pseudo random algorithm.
Note: When you set the Noise amplitude please pay attention that Signal
+ Noise does not exceed the amplitude of 6Vpp.
Use this control to enable or disable the noise added to the output signal.
This control allows you to invert the polarity of the output signal.
Sets the maximum voltage that the channel generates. This limit is
verified during the generation.
The constraint High Limit > Low Limit must be met. This feature can be
used to ensure the load is not damaged.
Note: The High Limit doesn't force setting voltages under the limit, but
during the generation the part of the waveform that exceeds the limit will
be cut at the High Limit.
Sets the minimum voltage that the channel generates. This limit is verified
during the generation
The constraint Low Limit < High Limit must be met.
This feature can be used to ensure the load is not damaged.
Note: The Low Limit doesn't force setting voltages under the limit, but
during the generation the part of the waveform that exceeds the limit will
be cut at the Low Limit.