Editing Programs in a Normal Sequence
To configure a program, select one of four operation modes (CC, CR, CV, or CP). You can
execute the same program repeatedly. You can also continue program execution by linking a
program to another program. Linked programs can only be executed if they have the same
operation mode and range. You can save up to 10 programs (No.1 to 10).
Press EDIT (SHIFT+A) to display a screen for configuring the programs in the sequence.
In the top screen, if you press
the bottom screen, if you press
In the screens for configuring the programs in the sequence, you can configure the following
items.
After you finish configuring the program settings, press NEXT (SHIFT+ ) to configure the
step settings.
Item
No.
Memo
Mode
Range
Loop
Last Load
Last Set
Chain
PLZ-4WH
SEQ
No.
: 1
Memo : Pro
g
ram1
Mode : NCC
Ran
g
e : 8.25A, 650V
Loop : 0001
PREV
NEXT
Top
Setting
1 to 10
Up to 11 characters
NCC: CC mode
NCR: CR mode
NCV: CV mode
NCP: CP mode
8.25 A, 650 V (example)
1 to 9998
9999: Infinite repetition
OFF: Load off
ON: Load on
0 to 100 % of the range
OFF: The sequence ends
1 to 10
when the cursor is at Loop, the bottom screen appears. In
when the cursor is at Last Load, the top screen appears.
SEQ
Last Load : ON
Last Set : 0.0000
A
Chain
: 2
PREV
NEXT
Bottom
Description
Program number (11 is for the fast sequence)
A memo
The program operation mode
The current and voltage ranges for the program
vary depending on the model.
The number of times that the program will repeat.
The state of the load (on or off) after the sequence
finishes.
The fundamental setting at the end of the
sequence.
This setting is valid when Chain is set to OFF.
Specifies the program that will be executed next.
Only programs with the same operation mode and
range can be executed.
4