Connecting to the Load Input Terminals
24
Connection procedure
1
Turn the POWER switch off.
2
Make sure that the output of the DUT is off.
3
Attach the lock plates to the load input terminals on the rear panel.
After you attach the lock plates once, there is no need to remove them.
4
Attach crimping terminals to the load wiring.
The DC INPUT terminals on the rear panel have open bolt holes (M8) for connecting
the load wiring. Attach appropriate crimping terminals.
5
Pass the wiring that you want to connect to the load through the
attached load input terminal cover.
6
Connect the load wires to the load input terminals on the rear panel
using the included load input terminal screw set.
Bolt (M8 × 18)
Spring washer (M8)
Insert the lock plates into
the slots by the terminals.
Lock plate
Crimping terminal
Load input terminal cover
Be sure to use the
included screw set.
Nut (M8)
PLZ-4WH