Page 3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE PMC PROCEDURE, NARKOMED GS The procedures in this section shall be performed in their entirety each time a component is removed, replaced, calibrated, adjusted and during all scheduled Periodic Manufacturer’s Certification (PMC) visits.
Page 4
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) Materials Required (continued): -- Fresh Gas Leak Test Adapter (P/N 4115041 or equivalent) -- Volumeter/Fresh Gas Adapter (P/N 4115042) -- Test Terminal 2x (P/N 4104389 or equivalent) -- Breathing System Leak Test Device (P/N S010159 or equivalent) -- PDM/Suction Adapter (P/N 4115038) -- Scavenger Adapter (P/N 4108114)
Page 5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 4114807 PRESSURE TEST ASSEMBLY , WITH ADAPTERS 4114830-002 4114830-001 4114830-004 4114830-003 SP15001 4114830-006 4114830-005...
Page 6
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) Liter vol/min S000081 2212300 FLOW METER MINUTE TEST STAND VOLUMETER 4104389 TEST TERMINAL ADAPTER 8401892 SIEMENS TEST LUNG SP15002...
Page 7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS S010158 FRESH GAS OUTLET VOLUME TEST DEVICE 4115041 FRESH GAS LEAK TEST DEVICE S010159 BREATHING SYSTEM LEAK TEST DEVICE 4115042 VOLUMETER/ FRESH GAS HOSE ADAPTER SP15003...
Page 8
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 4115038 PDM TO PATIENT SUCTION ADAPTER 4116111-001 4108114 NIBP W/LUER SCAVENGER ADAPTER TEST ADAPTER 4110709 LUER (F) 1/8 MPT ADAPTER FOR TOP PORT ON CAPNOMED FLOW METER 4115043 PDM TO...
Page 9
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS Periodic Manufacturer's Certification General Instructions The purpose of this manual is to provide detailed instructions for performing a Periodic Manufacturer’s Certification (PMC) inspection on a Narkomed GS Anesthesia machine. A PMC consists of a complete Periodic Manufacturer’s Service procedure and a certification level inspection based on Draeger Medical, Inc.
Page 10
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) Roman Numerals I and II do not affect the certification level but rather are provided to give further instructions to the end user as follows: I = The system in its present configuration shall only be used with a CO2 monitor incorporating an apnea warning.
Page 11
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS PM Certification Procedure for Narkomed GS Anesthesia System Use the PM Certification form for Narkomed 2B/ 2C/ GS Anesthesia Systems (P/N S010211). Completely fill in the header information. All Narkomed GS are equipped with Humphrey valves and do not require lubrication.
Page 12
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) Proceed with PM Certification procedure. If any tests fail refer to the “Failure Codes” listing in DMI Recommendations Guidelines Index (P/N S010250) to determine correct certification level starting point.
Page 13
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS SELF-DIAGNOSTICS 6.1.1 Turn the System Power switch to ON and verify the “ON” LED is lighted? 6.1.2 Verify all LED’s on the keypad and ventilator displays are lit if applicable. 6.1.3 Verify that the following is displayed on the alarm CRT: NARKOMED GS...
Page 14
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.2.1.4 Set the safety analyzer GROUND switch to NORMAL. Set the POLARITY switch to OFF. 6.2.1.5 Plug the machine’s AC power cord into the safety analyzer. 6.2.1.6 Apply the analyzer’s test lead to a cylinder yoke bolt.
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.2.4 Convenience Receptacle and Auxiliary Outlet Strip NOTE: This test will check the convenience receptacle and the auxiliary strip outlets for fault conditions such as open ground, reverse polarity, open line and open neutral.
Page 16
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) CALIBRATIONS - One Year Service Interval; Due Date _____ 6.5.1 To bring up the Oxygen Monitor Service Screen, press the Mon Cal key. 6.5.2 Remove the Oxygen sensor from the valve dome adapter, and remove the Oxygen sensor capsule from the Oxygen sensor housing.
Page 17
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.6.8.1 Remove the Ultrasonic Flow Sensor connector hose. 6.6.8.2 Is the connector hose, connector, and O-ring in good condition? ___ (Y) 6.6.8.3 Remove the expiratory valve. 6.6.8.4 Is the washer under the valve in good condition? ___ (Y) 6.6.8.5...
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.6.13 Are there two (2) spring clips on the absorber rods? ___ (Y) 6.6.14 Inspect the following: canisters, canister gaskets, dust cup and O-ring, and soda lime.
Page 19
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.8.1.5 Attach a test terminal (P/N 4104389) to the Fresh Gas Leak Test Adapter (P/N 4115041) then attach the test terminal to the bag mount. 6.8.1.6 Apply 50 cm H O test pressure to the absorber system and...
Page 20
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.8.3.2 Disconnect the hose from the expiratory valve or expiratory hose terminal on the ultrasonic flow sensor, if applicable and attach it to the test volumeter adapter. 6.8.3.3 Close the APL valve.
Page 21
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.8.5.9 Remove all test equipment. 6.8.6 PEEP Valve w/Bypass - If applicable 6.8.6.1 Open the APL valve. 6.8.6.2 Interconnect the inspiratory valve and expiratory valve or expiratory port on the ultrasonic flow sensor, if applicable with a 22 mm hose (P/N 9995132).
Page 22
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.8.7.5 Disconnect the pressure pilot line from the Bain Circuit and replace it with a PDM To Monitor Adapter (P/N 4115041). 6.8.7.6 Connect a test gauge to the adapter. 6.8.7.7 Occlude the expiration port on the Bain Circuit.
Page 23
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.9.14 After 1 minute, is the O concentration 97 to 100%? ___ (Y) 6.10 FLOWMETERS CONCENTRATIONS 6.10.1 Oxygen Flowmeter 6.10.1.1 Is it possible to adjust the flow of Oxygen over the full range of the flowmeters? ___ (Y) 6.10.1.2 Set the Oxygen flow to 4 l/min.
Page 24
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.10.6 Air Flowmeter - If applicable 6.10.6.1 If not configured with an Air Cylinder yoke, attach the Air Pipeline hose. 6.10.6.2 Is it possible to adjust the flow of the Air over the full range of the flowmeter? ___ (Y) 6.10.6.3 Set the Air flow to 2 l/min.
Page 25
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.10.10.2 Increase the pressure to 50 cm H 6.10.10.3 After 10 seconds, is the pressure within 40 to 60 cm H ___ (Y) 6.10.10.4 Remove the test gauge and adapter.
Page 26
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.11.2 Air Regulator - If applicable NOTE: Minimum cylinder pressure for Air regulator test is 1000 psi. 6.11.2.1 Configure the test gauge (P/N 4114807) using an Air nut/ stem DISS connector (P/N 4114830-002) on the hose and a DISS body connector (P/N 4114830-001) on the valve body side.
Page 27
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.11.3 Regulator NOTE: Minimum cylinder pressure for O regulator test is 1000 psi. 6.11.3.1 Configure a test gauge (P/N 4114807) using an O nut/stem DISS connector (P/N 4114830-006) on the hose and an O DISS body connector (P/N 4114830-005) on the valve body...
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.12.6 Remove the test equipment. 6.13 OXYGEN SUPPLY FAILURE PROTECTION 6.13.1 Connect all pipeline supplies. 6.13.2 Close the Oxygen flow control valve if applicable. 6.13.3 *Is the flow of Oxygen 150 to 200ml/min.? ___ (Y) 6.13.4...
Page 29
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.14.10 Increase the test pressure slowly. Does a VENT PRESS HI activate as a warning alarm within 47 to 53 cm H O? ___ (Y) 6.14.11 Create a sub-atmospheric test pressure slowly.
Page 30
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.15.18 Adjust the Inspiratory Flow until a peak pressure of 80 cm H O is achieved. 6.15.19 If applicable, set the Pressure Limit Control to within the 30 range. Readjust within the band as necessary to achieve proper value.
Page 31
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.16.2 Does the bellows remain fully inflated during the expiratory pause phase? ___ (Y) 6.16.3 Remove the ventilator hose from the VENTILATOR HOSE terminal. 6.16.4 Attach a test terminal to the bellows assembly ventilator hose terminal.
Page 32
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.17.6 Adjust the O flow to 3 l/min. 6.17.7 Is the tidal volume on the test volumeter greater than 250 ml? ___(Y) 6.17.8 Adjust the tidal volume to the 100 ml mark on the pediatric bellows assembly.
Page 33
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.18.9 Adjust the Oxygen flow to 10 l/min. 6.18.10 Is the PEEP displayed on the monitor within 0 to 3 cm H O? ___ (Y) 6.18.11 Close the Oxygen flow control valve.
Page 34
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) Go to step 6.19.2.5. 6.19.1.4 Cleaning procedure for assemblies with many small vent ports. Remove the scavenger mounting screws. Remove the scavenger flow control needle valve assembly.
Page 35
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.19.2.12 What is the pressure on the test gauge? ___ cm H O (0 to -0.5) 6.19.2.13 Remove the test equipment, re-cap the scavenger adapter port or remove the scavenger adapter and reconnect the scavenger hose.
Page 36
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.20.2.7 Connect a test pressure monitor to the hose barb on the adapter using a PDM/Suction Adapter (P/N 4115038). 6.20.2.8 Remove the transfer hose from the bottom of the scavenger and occlude this port.
Page 37
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.21.1.10 Reinstall the valve in the housing, and then reinstall the housing into the scavenger body. NOTE: This valve does not require washers or O-rings. 6.21.1.11 Remove the valve housing on the right side of the scavenger body by turning it counter-clockwise.
Page 38
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS NMGS PMC PROCEDURE (continued) 6.21.2.15 Remove the test equipment and reconnect the scavenger hose. 6.21.2.16 Close the Oxygen flow control valve. 6.22 SUCTION REGULATOR - 6-month Service Interval; Due Date _____, If applicable 6.22.1 Set the vacuum on/off valve to the ON position.
Page 39
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMC PROCEDURE (continued) NMGS 6.24.5 Verify the APL valve knob is turned completely counterclockwise (fully open). 6.24.6 Place the AUTO/BAG selector in the BAG position. 6.24.7 Verify the ventilator hose is connected between the Auto/Man valve and Ventilator hose terminal.
Page 40
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 41
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 42
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS DrägerService is a division of Draeger Medical, Inc. 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047 Fax: (215) 721-5784 Web: www.draegermedical.com Printed in the U.S.A.