OPTISONIC 6300
3.4.2 Installation requirements signal converter
• Allow 10...20 cm / 3.9...7.9" of space at the sides and rear of the signal converter to permit
free air circulation.
• Protect signal converter against direct solar radiation, install a sunshield if necessary.
• Signal converters installed in switchgear cabinets require adequate cooling, e.g. by fan or
heat exchanger.
• Do not expose the signal converter to intense vibration.
INFORMATION!
For detailed information please also refer to Housing on page 115
3.5 Installation requirements sensor
INFORMATION!
To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an
empty pipe, please observe the following precautions.
CAUTION!
Since gas will collect at the highest point of a pipe, installation of the flowmeter at that location
should be avoided at all times. Also installation in a down going pipe should be avoided since a
completely filled pipe may not be guaranteed due to cascading effects. Additionally flow profile
distortion is possible.
CAUTION!
If you program the diameter, please note that you use the outer diameter of the pipe.
07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en
www.krohne.com
INSTALLATION
.
3
17