Installing The Device; Post-Installation Check - Endress+Hauser Liquipoint FTW33 Operating Instructions Manual

Conductive and capacitance point level measurement
Table of Contents

Advertisement

Liquipoint FTW33
1)
Can be ordered as an optional accessory
Endress+Hauser
 5
Flush-mounted installation
5.1.3
Special mounting instructions
• Protect housing against impact.
• Moisture must not penetrate the housing when mounting the device, establishing the
electrical connection and during operation.
• In the IP69 version, only remove the protection cap from the M12 plug shortly before
establishing the electrical connection.
5.2

Installing the device

5.2.1
Required tools
Open-ended wrench or, for measuring points that are difficult to access, a hexagon tubular
1)
socket wrench 32 mm
– When screwing in, turn by the hex bolt only.
– Torque: 15 to 30 Nm (11 to 22 lbf ft).
5.2.2
Installation
A
32 mm
A
Thread G ½"
B
Thread G ¾"/G 1"
C
Thread M24 × 1.5
5.3

Post-installation check

 Is the device undamaged (visual inspection)?
 Does the device comply with the measuring point specifications?
– Process temperature
– Process pressure
– Ambient temperature range
– Measuring range
B
32 mm
Installation
A0025915
C
32 mm
A0021389
13