Diagnosis - Endress+Hauser Liquipoint FTW33 Operating Instructions Manual

Conductive and capacitance point level measurement
Table of Contents

Advertisement

Description of Device Parameters
Actual Diagnostics (STA)
Navigation
Description
Last Diagnostic (LST)
Navigation
Description
Simulation Switch Output 1 (OU1)
Navigation
Description
Options
Simulation Switch Output 2 (OU2)
Navigation
Description
Options
34
15.2

Diagnosis

Diagnosis → Actual Diagnostics (STA)
Displays the current device status.
Diagnosis → Last Diagnostic (LST)
Displays the latest device status (error or warning) that was rectified during operation.
Diagnosis → Simulation Switch Output 1 (OU1)
The simulation affects the process data only. It does not affect the physical switch output.
If a simulation is active, a warning to this effect is displayed so that it is obvious to the user
that the device is in simulation mode. A warning is communicated via IO-Link (C485 - sim-
ulation active). The simulation must be ended actively via the menu. If the device is discon-
nected from the power supply during the simulation and power is then resupplied,
simulation mode is not resumed, and instead the device continues operation in measuring
mode.
• OFF
• OU1 = HIGH
• OU1= LOW
Diagnosis → Simulation Switch Output 2 (OU2)
The simulation affects the process data and the physical switch output. If a simulation is
active, a warning to this effect is displayed via IO-Link so that it is obvious to the user that
the device is in simulation mode (C485 - simulation active). The simulation must be ended
actively via the menu. If the device is disconnected from the power supply during the simu-
lation and power is then resupplied, simulation mode is not resumed, and instead the
device continues operation in measuring mode.
• OFF
• OU2 = HIGH
• OU2 = LOW
Liquipoint FTW33