Endress+Hauser Liquipoint FTW33 Operating Instructions Manual page 35

Conductive and capacitance point level measurement
Table of Contents

Advertisement

Liquipoint FTW33
Device search
Navigation
Description
Note
Options
Factory setting
Sensor check
Navigation
Description
Options
Endress+Hauser
Diagnosis → Device search
This parameter is used to uniquely identify the device during installation.
The green LED is lit (= operational) on the device and starts to flash with increased lumi-
nosity, flash frequency
There is no external signaling via LEDs on the metal housing cover (IP69).
• OFF
• ON
OFF
Diagnosis → Sensor check
This parameter is used to test if the measuring point is functioning correctly.
The sensor must not be covered and must be free of residue. The device compares the cur-
rent measured values with the measured values from the factory calibration.
The device must be removed before the sensor check since the free value is influ-
enced by the type of installation.
Check: Following the test, one of the following messages is displayed:
• Message (0x1814) for sensor check passed
• Message C103 (0x1813) for sensor check failed
Description of Device Parameters
.