1.0 Introduction

1.2 Function

The ATMOS A / C 161 / 261 Aspirator is an exceptionally
handy small suction device.
It is driven by an electromotive, maintenance-free swing
piston pump. During operation the pump generates a
vacuum within the hose system and the secretion canister
by means of which secretion and air can be sucked off
(using a sterile suction catheter).
The fluid is collected in the secretion canister.
ATMOS A 161 / 261 Aspirator
The final vacuum and thus the suction capacity can be
adjusted by means of the three step adjustment.
An overtemperature stop avoids overheating of the suction
device.
ATMOS C 161 / C 261 Aspirator
The final vacuum and thus the suction capacity can be
adjusted by means of the fine regulation and the vacuum
gauge.
An overtemperature stop avoids overheating of the suction
device.
ATMOS A / C 161 / 261 Aspirator / DDS
The reusable secretion canister is connected to the pump
housing via direct-docking, without any pedestrian hose
system. Now the user can/must only connect the suction
hose.
A DDS bacterial and viral filter located in the canister lid
prevents bacteria, viruses and liquids from entering the
device.
A mechanical overflow protection (float ball) is integrated in
the canister lid. This prevents an accidental absorption of
secretion into the pump head.

1.3 Intended use

Name: ATMOS A 161 Aspirator / ATMOS A 261 Aspirator
ATMOS C 161 Aspirator / ATMOS C 261 Aspirator
Main function: Temporarily and spontaneously suction of
secretions, blood and body fluids which typically accumulate
during respiratory tract suction and ENT treatment.
Med. indications / application: Suction of the upper
respiratory tract and throat, nose ear.
Specification of the main function: Drainage and
temporary collection of body fluids. An electric suction
pump is used to generate negative pressure. The integrated
suction canister allows a temporarily collection of the
derived body fluids.
Application organ: Upper respiratory tract (oral cavity,
nasopharyngeal cavity and bronchial system) and throat,
nose, ear.
Application time: Temporary use on the patient during
respiratory tract suction (< 60 min.) or short time use at ENT
treatment (up to 30 days).
Application site: The application site of respiratory tract
suction is the clinical, practice, nursing and home care
sector. In ENT medicine the devices are used in clinics
and practices. The application of the device may only be
performed by medically trained and introduced staff .
4
!
Contraindications:Not suitable for:
• continuous operation for drainage procedures in the low
vacuum range (e.g. cardiothoracic or wound drainage)
• permanent endoscopic use
• the use outside the medical sector
• suction of flammable, corrosive or explosive substances
• suction in potentially explosive areas
The product is:
Sterility: The device is not sterile.
Single-use product/reprocessing: The product is intended
for multiple use. The device and part of the accessories are
reusable. For information on reprocessing, cleaning and
disinfection please see the operating instructions.
Fig. 1a. ATMOS A / C 161 / 261 Aspirator / DDS
 Sliding cover
 Vacuum regulator (individually adjustable)
 Vacuum gauge
 Hose holder
Fig. 1b. ATMOS A / C 161 / 261 Aspirator / DDS
 Vacuum regulator (3 steps adjustable)
 Hose holder
X active
□ not active