14.0 Notes on EMC
Guidelines and Manufacturer's Declaration - Immunity for
ATMOS A / C 161 / 261 Aspirator
The ATMOS A / C 161 / 261 Aspirator is designed for operation in the electromagnetic environment specified below. The
customer or user of the ATMOS A / C 161 / 261 Aspirator should ensure that it is used in such an environment.
Immunity Test
ESD acc. to
IEC 61000-4-2
EFT IEC 61000-4-4
Surges IEC 61000-4-5
Voltage Dips / Drop out
IEC 61000-4-11
Magnetic field at power
frequency
(50/60 Hz) acc. to IEC
61000-4-8
NOTE UT is the alternating mains voltage prior to application of the test levels.
24
IEC 60601- Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
± 1 kV common mode
± 2 kV differential mode
< 5 % UT
(> 95 % Dip of the UT for 0,5
cycles)
40 % UT
(60% Dip of the UT for 5
cycles)
70 % UT
(30% Dip of the UT for 25
cycles)
< 5 % UT
(> 95 % Dip of the UT for 5 s)
3 A/m
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV common mode
± 2 kV differential mode
< 5 % UT
(> 95 % Dip of the UT for 0,5
cycles)
40 % UT
(60% Dip of the UT for 5
cycles)
70 % UT
(30% Dip of the UT for 25
cycles)
< 5 % UT
(> 95 % Dip of the UT for 5 s)
Not applicable!
Electromagnetic Environment
– Guidance
Floors should be made of wood
or concrete or tiled with ceramic
tiles. If floors are synthetic, the
relative humidity should be at
least 30 %.
The quality of the supply
voltage should correspond to a
typical commercial or hospital
environment.
The quality of the supply
voltage should correspond to a
typical commercial or hospital
environment.
The quality of the supply
voltage should correspond to a
typical commercial or hospital
environment. If the user of
the ATMOS A / C 161 / 261
Aspirator requires continued
operation even when power
supply interruptions occur, it is
recommended that the ATMOS
A / C 161 / 261 Aspirator be
powered by an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.