How to Operate the Blanket Warmer
Before you begin
Procedure
During the warming
process
Result
B la n k et W ar m er
Op e r at or ' s Man u a l
▪
The warmer must be connected to electric power and turned on.
WARNING:
Do not operate the warmer in the presence of flammable
anesthetic mixtures (with air or with oxygen or nitrous oxide).
WARNING:
Make sure the blanket support assembly and shelves are
installed to prevent the blankets from being scorched or
discolored.
Do not overload the cavity.
NOTICE
Allow 1" (25mm) between the blankets and the interior walls.
To operate the warmer, do the following.
Step
Action
1.
Set
the temperature using the arrow buttons
NOTE:
(32°C – 71°C).
Open
2.
the door and load the dry, cotton blankets into the warmer.
Close
the door.
NOTICE
3.
Press
the temperature recall button
temperature.
The measured cavity temperature will display for five seconds. Then, the
set-point temperature will display.
Rotate
4.
the blankets daily from the bottom to the top of the load to ensure
equal heating and to prevent discoloration.
The blankets are now warming.
MN -4 6 42 7
Re v 1
6 / 1 9
▪
▪
▪
Personal injury hazard.
Personal injury hazard.
The temperature set-point range is 90°F – 160°F
Do not warm items containing plastic, rubber, metal
snaps, studs, hooks, etc.
to view the measured cavity
O
PERATION
.