Page 1
Operation and Care Manual EC Series Blanket Warmer EC1540 EC2060 120v EC1540 Shown with Optional EC2060 Full-Perimeter Rubber Bumper Enthermics Medical Systems An ISO 13485:2003 certified company PO Box 443, Menomonee Falls WI 53052-0443 W164 N9221 Water St, Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics.com...
Appliances on casters can move or roll on uneven floors. leveling appliance and handling accessories. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
NOTICE: For equipment delivered for use in any location regulated by the following directive: 2012/95/EC WEEE Do not dispose of electrical or electronic equipment with other municipal waste. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
The appliance is furnished with one (1) set of an additional safety feature. 5" (127mm) heavy-duty casters – two (2) rigid and two (2) The controller displays the temperature in whole degrees. swivel with locking brake. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Laboratories with Respect to Electrical Shock, UL File No. Fire and Mechanical Hazards only, in Accordance E201645 Protective Earth with UL 60601-1 and CAN/CSA C22.2 No. 601.1. Ground Symbol MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
2. Make sure the product inside the appliance is at the Temperature correct operational temperature before using. 1. Press and hold the "TIMER" button and the up arrow button simultaneously. The temperature at the compartment sensor is displayed MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 11
The timer status indicator illuminates. How to Disable the Timer 1. Press and hold the "TIMER" button until the timer status indicator goes off. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 12
NOTE: Failure to rotate the blankets can cause blankets to discolor a er an extended period of time. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
7. Clean the appliance glass with glass cleaner or minimum of once per day. Always wipe standing water as distilled vinegar. quickly as possible. 8. Replace the support assembly. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 14
Do not allow the thermocouple to touch any surface. Monitor the temperature for one hour in an empty interior. NOTE: The blanket warmer temperature may fluctuate ±10° from the set point. *Contact service for immediate repair MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Part numbers and drawings are subject to change without notice. • Replace the defective relay. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer • If the error continues, call service. E-50 Temperature...
Page 16
Part numbers and drawings are subject to change without notice. E-87 EEPROM Error • Ensure that all calibration offsets are properly set. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer • If the error continues, call service for proper calibration instructions. E-90 A controller button...
Service EC1540 Full Assembly and Interior Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 18
Service EC2060 Full Assembly and Interior Blanket Insert Assembly Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 19
GS-38504 * Not shown **Quantity varies † Quantity doubles for pass-through option Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 20
Service EC1540 Electrical EC2060 Electrical Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 21
24 Wire diagram* (located under top cover) 7756 77563 * Not shown † Optional Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 22
SL-3063 Insulating sleeve CR-3226 Ring connector BU-3105 Shoulder bushing ST-2439 Stud, 10-32 Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Wiring Diagram Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 24
Wire Diagram Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-28667 • Rev 4 • 12/16 • EC Series Blanket Warmer...
Page 25
Enthermics Medical Systems An ISO 13485:2003 certified company PO Box 443, Menomonee Falls WI 53052-0443 W164 N9221 Water St, Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics.com www.enthermics.com Printed in the U.S.A. Specifications are Subject to Change Without Notice...