Maintenance Schedule
Daily
Monthly
Yearly
Bl a n k et W ar m er
Op e r at or ' s Man u a l
▪
For daily maintenance, do the following.
Check:
▪
the air vents in the airflow insert panels are not obstructed (if applicable).
▪
all fan guards are clear and not obstructed (if applicable).
▪
the number of bottles/bags, or blankets as applicable do not exceed the
maximum capacity per shelf or basket.
Clean:
▪
any spills with a clean, lint free cloth.
For monthly maintenance, do the following.
Check:
▪
the door gasket for tears and holes. Make sure that it is firmly attached to the
door. Check the seal when the door is closed.
▪
the guards around the air temperature sensor are in place and fully secure to the
warmer.
▪
the hardware securing the warmer(s) to the wall, if applicable.
Clean:
▪
the outside of the warmer with stainless steel cleaner.
▪
vacuum (if applicable):
fan openings
□
fan sail switch
□
vent openings
□
For yearly maintenance, do the following.
Check:
▪
the set-point temperature compared to the actual temperature displayed.
Check the cavity air temperature with a thermocouple placed 1" (25mm) from
□
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
▪
the condition of the plug and cord and replace if damaged.
▪
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
▪
the shelves and interior of the warmer.