Installation
7
8
9
Once completed, the date and time of the last successful calibration will be
NOTE
displayed, a pass or fail mark will appear along with the wavelength error indicating
whether the instrument calibration is complete.

Saving and viewing the calibration data

The calibration data is stored in the Logs. To view the data, export the Logs.
1
2
3
The default save location is \Users\'Username'\Documents\Agilent\ICP Expert
NOTE
\Exported Results\ on the drive on which the ICP Expert software is installed.

Hardware Components Replacement

Your ICP-OES should be ready to operate after the Agilent field service engineer
has installed it. However, you may need to set up items such as the peristaltic
pump, nebulizer, torch or spray chamber. For instructions on how to do this, please
refer to the 'How to' section in the ICP Expert Help and Learning Center.
Agilent 5800 and 5900 ICP-OES User's Guide
Check that the Polychromator Boost purge is on.
a
On the 'Instrument' window of the ICP Expert software, click the Status tab
and check that under the Polychromator section Boost is selected.
b
If the instrument was turned on from an idle state, the wavelength
calibration in Steps 7-10 can be performed immediately. If the instrument
was turned on after more than a few hours of being unpowered, the
polychromator can take several hours to stabilize the temperature and to
purge.
Aspirate the wavelength calibration solution and then select the Calibration
tab.
Click Calibrate in the 'Instrument' section to perform a wavelength calibration.
From the main ICP Expert window, click File > Logs > Export Logs.
Save the zipped Log file.
Open the .zip file to view the Log files.