Maintenance and Troubleshooting
2
3
Long term exposure to acid during cleaning may lead to discoloration of the plastic
NOTE
base. This change is cosmetic only and should not impact performance if the torch
is clean and the results of the other torch checks are satisfactory.
Store the torch in the original box or a plastic bag when not in use, or alternately,
TIP
use the Agilent ICP-OES Easy-fit torch storage rack (P/N G8010-67000).

Troubleshooting

For troubleshooting information, please see the ICP Expert Help and Learning
Center:
1
2

Spare Parts

For spare parts and consumables ordering information, refer to the Agilent
Technologies website:
www.agilent.com
To replace the items listed below, you must use Agilent-manufactured parts, which
can be ordered online from the Agilent website or through your local sales
representative.
Agilent 5800 and 5900 ICP-OES User's Guide
Check for carry-over after refitting the torch to the instrument to determine if
the cleaning procedure has been sufficient. If carry-over is found, repeat the
cleaning process.
Replace the torch when the outer surface of the quartz outer tube is rough to
the touch (which indicates signs of wear), or if there are any cracks visible.
Double-click the ICP Expert Help and Learning Center on the computer
desktop.
Click on, or search for Troubleshooting to view instructions on how to
troubleshoot.