Maintenance and Troubleshooting
Annually

Cleaning

Any spills in the sample compartment should be wiped up immediately.
The user (or other authorized personnel) must perform the appropriate
decontamination procedure if hazardous material is spilled on or inside the ICP-OES.
The exterior surfaces of the ICP-OES should be kept clean. All cleaning should be
done with a soft cloth. If necessary, this cloth can be dampened with water or a
mild detergent. Do not use organic solvents or abrasive cleaning agents.
Before using any cleaning agent, procedure or decontamination method except
those specified by Agilent, the user (or other authorized personnel) should check
with your local authorized Agilent field service engineer or representative to
confirm that the proposed method will not damage the equipment.

Torch Cleaning

To help you achieve the maximum usable life from all models of the 5800/5900
ICP-OES Easy-fit torch, it is recommended that the cleaning procedures are
followed as soon as any discoloration appears on the outer tube of the torch.
To maximize torch lifetime and prevent contamination, you should inspect the
condition and cleanliness of the torch daily.
See the ICP Expert Help for instructions on how to clean the torch for organic
based samples and have carbon build-up on the injector.
54
Drain and replace chiller coolant.
Replace the argon purge gas filter.
Remove and clean the water filter on the right side of the instrument.
Periodically, drain the coolant from the cooling system and then refill/ treat
with an appropriate algaecide (as recommended by the manufacturer).
Agilent 5800 and 5900 ICP-OES User's Guide