Installation

Assembly of the fully demountable torch

To assemble the torch:
1
2
3
4
5
Hand tighten only. Do not overtighten or use a wrench for leverage.
CAUTION
Agilent 5800 and 5900 ICP-OES User's Guide
Figure 8. Ensure the seal is properly seated by pressing firmly on the top edges
Unscrew the locking nut on the torch base.
Place the top seal back on the top of the torch base and ensure this is properly
seated by pressing firmly on the top edges.
Insert the tube-set into the torch body and position it so that the groove on the
torch body aligns with the "T" alignment mark on the tube-set. Press the outer
tube-set fully into the torch body. See Figure 7 and Figure 8.
Fully insert the injector through the base of the torch body until the bulge
contacts the retaining arm and "clicks" into position.
2
Replace the locking nut and tighten to secure the injector.
1
31