What to do if...?
13 What to do if...?
If electrically operated functions do not work properly, always check these points first:
• Is the power switch switched on?
• Are the power cables attached correctly?
• Are all connecting cables attached correctly?
• Are all video cables attached correctly?
13.1 General
For malfunctions regarding the Leica surgical microscope refer to the user manual of the Leica surgical microscope.
13.2 GLOW800
13.2.1 Calibration
Observation
The GLOW800 injection in the eyepiece does not
fit with the optical image.
The GLOW800 fluorescence image is incorrectly
aligned to the White Light image
13.2.2 VPU
Observation
Incorrect information is shown for interpretation
to the user
The system is not booting up
The video signal shows red cross /Red "X"
The system is frozen/ no reaction
13.2.3 Limitation
Observation
In high magnification the GLOW800 fluorescence
is out of focus, although the White Light image is
sharp
32
Cause
The CaptiView calibration is lost
The setting to fit both images are incorrect
Cause
Device algorithm calibration failure
Improper pre-operational check
Electronic failure
Camera signal lost
Software initialization failure
Cause
In high magnification the NIR focus can differ
from the WL focus
GLOW800 / Ref. 10 747 241 / Version 04
Remedy
The calibration has to be redone.
X
X
Get in contact with Leica Service.
X
X
The settings have to be adapted to fit both
X
X
images.
Get in contact with Leica Service.
X
X
Remedy
Use reset button to reboot/reset the system.
X
X
Repeat pre-operational procedure.
X
X
Get in contact with Leica Service.
X
X
Get in contact with Leica Service.
X
X
Use reset button to reboot/reset of the
X
X
system.
Get in contact with Leica Service.
X
X
Use reset button to reboot/reset the system.
X
X
Remedy
Reduce the magnification until your receive a
X
X
sharp GLOW800 image