Preparation before surgery
7.6

GLOW800 image injection

With the optional CaptiView image injection integrated into the
surgical microscope M530 ARveo, OH6 and OHX, the GLOW800
fluorescence information can be observed in the right, left or both
eyepieces, as an injected digital image or as an overlay, according to
the selected fluorescence observation type and the CaptiView
settings.
To decide about observation type A or B see chapter 7.3.1, to decide
about observation types A-1 to A-4, means the CaptiView user
settings see chapter 7.7.
Observation Type A-1/WLFL
Digital white light image with embedded fluorescence in
pseudocolor video in the right eyepiece and optical 2D image in the
left eyepiece
Observation Type A-2/OVR
Digital pseudocolor fluorescence image overlay in the right
eyepiece to the stereoscopic observation of the optical images in
both eyepieces
For correct fitting fluorescence overlays in the eyepiece(s) it is recommended to check the fluorescence image focus on the monitor.
To avoid a double recorded or displayed fluorescence images on the monitor switch off the overlay in the left eyepiece.
22
Observation Type A-3/OVL
Digital pseudocolor fluorescence image overlay in the left eyepiece
to the stereoscopic observation of the optical images in both
eyepieces
Observation Type A-4/OVLR
Digital pseudocolor fluorescence image overlay in both eyepieces
to the stereoscopic observation of the optical images in both
eyepieces
GLOW800 / Ref. 10 747 241 / Version 04