8.3

Preparation

WARNING
User selected with a programmed fluorescence function.
Correct user is activated.
X
X
Preparational check is performed.
X
X
Microscope illumination lamp is within tolerances (see
X
X
user manual of the Leica surgical microscope).
Please prepare the test as follows:
For general tests please use the GLOW800 preset.
X
X
Place the GLOW800 test card below the microscope.
X
X
WARNING
Risk of infection due to insterile GLOW800 test card.
Do not use the GLOW800 test card in the sterile field.
X
X
Use only in non-sterile environment.
X
X
Check the microscope illumination in non-sterile
X
X
environment only.
Take care to ensure the precise parfocal setting of the
X
X
Leica surgical microscope. Follow the instructions on
parfocal setup.
Adjust the working distance (WD) to 350 mm.
X
X
To avoid reflexes position the microscope in a small, but
X
X
sufficient angle over the test card.
Follow the instructions on parfocal setup.
X
X
Get in focus by repositioning the microscope at highest
X
X
magnification (Do not focus!).
After positioning and focussing adjust the magnification to
X
X
3.0×.
Move the test card into the center of the field of view.
X
X
Switch to GLOW800 mode by pressing the GLOW800 On/Off
X
X
button on the handle.
Adjust the fluorescence illumination to 50 %.
X
X
The GLOW800 test card can now be observed in the eyepiece in
X
X
white light and on the optional monitor in white light
fluorescence.
1
1 Microscope optics carrier
2 Test card
Appearance of the test card in white light observation:
Holes in the colors squares will allow to check the adjustment of the
fluorescence and white light image.
GLOW800 / Ref. 10 747 241 / Version 04
Check the illumination and function and adjustment
2
Eyepiece view:
Leica GLOW800
10 746 513-xx
0
1
2
3
4
5
6
7
8
9
!
Please read the Leica
GLOW800 user manual!
25