Preparation before surgery
7.3.4
Intensity
«Intensity» bar (4, page 14) adjusts the fluorescence contrast,
brightness and transparency in relation to the object details.
The intensity values range from 0 % to 100 %, default is 50 %.
Fluorescence is visible across the full percentage range.
E.g. at 0 % intensity fluorescence is just visible with the object
dominant, at 100 % intensity fluorescence is intensively visible and
dominant.
Minimum intensity setting – 0 %
Medium intensity setting – 50 %
Maximum intensity setting – 100 %
7.3.5
WL Saturation and WL Brightness
To increase the contrast of the fluorescence information in relation
to the anatomical video image the characteristics of the white light
anatomical image can be adapted by:
• Reduction of the white light anatomical image color saturation
with the WL Saturation bar (5, page 14)
• Reduction of the white light anatomical image brightness with
the WL Brightness bar (6, page 14)
7.3.6
Apply button
By pressing the "Apply" button (7, page 14) on the GLOW800
18
page of the user settings changes of one or more of the different
GLOW800 settings become active in the GLOW800 system.
The result of the changed settings can be observed after a few
seconds on the GLOW800 screen.
7.4
7.4.1
With «Digital Zoom» the display format is adjusted to the needs of
each GLOW user. There are four formats available which can be
changed with the Digital Zoom button (3).
Default setting is "Max. Height Centered".
Max. Height
Fits the maximum detected field of view in the centre of the screen
of the documentation monitor.
• Nearly round image format – full video resolution
GLOW800 / Ref. 10 747 241 / Version 04

User settings camera

Digital zoom