9

Operation

WARNING
Danger of injury to the patient due to not approved
fluorescence media.
Only use fluorescence media approved for the planned
X
X
application.
When using GLOW800 please ensure to have a Doppler
Ultrasound or similar in place, in case of none or insufficient
blood flow visualization out of the ICG/GLOW800 procedure
is given.
9.1

Using the GLOW800

Switch on the illumination of the Leica surgical microscope.
X
X
Select a user: Select either the "Fluorescence Vascular
X
X
GLOW800" user preset or an own GLOW800 user.
9.2
Controlling the GLOW800
functions
Controlling the GLOW800 functions, e.g., on the microscope's left
handle
left handle
GLOW800
Play-
On/O
back
Step Loop
GLOW800 On/Off
The handle permits switching between white light mode
(GLOW800 On/Off) and GLOW800 mode.
Press the joystick to the left to switch between the modes.
X
X
WARNING
Danger of injury to the patient due to excessive GLOW800
radiation.
Avoid extended and/or excessively frequent use of
X
X
GLOW800 radiation.
right handle
Y+
X–
X+
Y–
GLOW800 / Ref. 10 747 241 / Version 04
GLOW800 mode is disabled automatically no later than after
180 seconds (preset, the value can be changed) to prevent
excessive exposure of the patient to GLOW800 radiation.
The "GLOW800 On/Off" function activates the GLOW800
illumination and the GLOW800 sensitive camera, connects
the GLOW800 video signals to the system's video output,
ports and the video is displayed on the monitor.
At the same time the recording of the GLOW800 video
signals is started at the optional connected HDMD PRO
recording unit.
When the "GLOW800 On/Off" button is pressed again, the
system returns to white light mode and disables the
GLOW800 functions and the GLOW800 recording is finished.
Replay
If an optional HDMD PRO recording unit is provided, pressing
X
X
the joystick to the right starts the replay of the last recorded
loop on the recording unit.
Step loop
If an optional HDMD PRO recording unit is provided, by
X
X
repeatedly pressing the joystick downward, you can switch back
to previously recorded GLOW800 loops.
Operation