Changing The Water Conversion Factor (Wcf); Operator Responsibilities - BWT AQA therm MOVE Power Installation And Operating Manual

Mobile reverse osmosis unit
Table of Contents

Advertisement

Available languages

Available languages

EN
4.
Press the circle button to start the rinsing
process. The rinsing process takes ap-
prox. 60 seconds. Following the rinsing
process, the production of outlet water
begins and the LED indicator remains lit
up in green.
5. Let the outlet water flow into a waste water
drain system for approx. 60 seconds until a
consistently low conductivity value (below 80
μS/cm) is set.
6. Then pause the rinsing process and connect
the outlet water hose to the system to be
filled.
7. Open the check valve positioned down-
stream from the product.
Reset and save the outlet water coun-
ter (Outlet Water Counter) in the app
to track the new capacity of the mem-
brane cartridges.
8. Resetting the outlet water counter in the
BWT RO systems app:
►Press and hold the checkmark signal and
then press the button again to confirm the
prompt "
RESET?
9. Restart the AQA therm MOVE Power.
The charging bar on the product display
shows the system pressure. When the set
switch-off point for pressure production is
reached, the pump switches off.
50
".
9 The system is filled.
9 You can read the fill level on the daily litres
counter on the display.
10 .1 Changing the water conversion
factor (WCF)
To change the water conversion fac-
tor, contact BWT customer service.
The factory setting for the water conversion fac-
tor (WCF) in the product is below 50% at a water
temperature of 15 °C. This minimises blockages
and increases the service life of the membrane
cartridges.
In some application scenarios, it may make sense
to increase the water conversion factor. For examp-
le, if there is a softening unit downstream from the
product in a stationary application, then you can as-
sume there will be minimal blocking of the membra-
ne cartridges and the water conversion factor can
be increased without any issues.
11

Operator responsibilities

Flawless function requires:
● The correct operation of the product.
● Operation or rinsing of the product at least once
every two weeks.
● Regular checks and servicing of the product.
Wearing parts must be replaced regularly within
the prescribed maintenance intervals.
● Before using the product, check with the wa-
ter supplier regarding the quality and pressure
conditions of the inlet water. If the water quali-
ty varies, you may have to change the product
settings.
● Check the product and connection lines for
damage before each use.
● Replace membrane cartridges if the conductiv-
ity value of the outlet water is ˃ 80 μS/cm or
every 3 years (see chapter „5.6 Fitting the new
membrane cartridges", Seite 37).
● Regular product checks by an electrician in ac-
cordance with DIN VDE 0701.