Other Observation Methods; Differential Interference Contrast Observation - Olympus BX61WI Instructions Manual

Fixed-stage motorized upright microscope
Table of Contents

Advertisement

OTHER OBSERVATION METHODS

5-1 Differential Interference Contrast Observation

# The normal optical performance of DIC observation cannot be manifested if a plastic petri dish is used.
}DIC prisms (for revolving nosepiece and condenser), an analyzer and a polarizer are required for DIC observation.
When the reflected light fluorescence illuminator is not used, the U-KPA intermediate tube is required to attach the analyzer.
4
1
2
1
1
24
3
Fig. 31
2
3
Fig. 32
Fig. 33
1
Attaching the Analyzer
U-ANT Analyzer
}Drop the U-ANT analyzer in the dummy slider of the U-KPA intermediate
tube.
1. Place the U-ANT 1 with the side with indications facing up, align the
indices and drop the analyzer into the dummy slider 2 (the analyzer will
be absorbed by magnet).
2. Set the dummy slider 2 back into the U-KPA 3 and tighten the clamping
knob 4.
U-AN Analyzer
}Insert the U-AN into the analyzer insertion slot of the reflected light fluo-
rescence illuminator. (Refer also to the instruction manual of your
reflected light fluorescence illuminator.)
2
Attaching the Polarizer
}The performance of a polarizer deteriorates after it has been subjected to
light for a lobe period (about 500 hours of continuous use). Replace the
polarizer when it has been used for a long period.
Drop the polarizer into the filter insertion position with a push ring 1 or 2,
and clamp with the push ring.
}It is recommended to insert the polarizer in insertion position 1. This is
because the polarizer rotation clamping knob 3 comes on the front of
the microscope when the insertion position 1 is engaged in the light
path.
}When the 32PO polarizer is used, adjustment is easier than with the
32POIR since images are brighter with the 32PO.
Besides, removal of the IR filter (32BP775 or 32IR900) makes images
brighter during adjustment although infrared light observation is to be
performed.
When Using the U-UCD8
# The polarizer built into the U-UCD8 is not necessary with this micro-
scope. Remove it as described below.
1. Using a Phillips precision screwdriver, remove the 6 clamping screws 1
retaining the polarizer cover at the bottom side of the condenser. (Fig. 33)
2. Remove the cover to expose the polarizer and remove it together with
the frame. (Retain the removed polarizer carefully for future use with a
microscope other than the BX61WI.
3. Attach the polarizer cover to the original position.
(Fig. 31)
(Figs. 32 & 33)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents