28 5. Operation
5.3.4 Selecting the calibration method (1P, 2P or 3P method)
1.
Open setting mode.
2.
Via MODE, go through the settings until CAL 1P, 2P OR 3P
appears.
3.
Select calibration method 1P, 2P OR 3P via CAL
4.
Press MODE to confirm
•
1P: can be selected via CAL (4), (7) or (10)
•
2P: can be selected via CAL (4,7) or (7,10)
5.
Press MODE to confirm selection
6.
Via MODE, go through the settings until all segments light up
briefly.
7.
The instrument then switches to the measuring mode.
5.3.5 Switching Auto Off on and off
1.
Open setting mode.
2.
Via MODE, go through the settings until AUTO OFF appears.
3.
Select On or OFF via CAL
4.
Press MODE to confirm selection
5.
Via MODE, go through the settings until all segments light up
briefly.
6.
The instrument then switches to the measuring mode.
5.3.6 Switching the beeper on / off
1.
Open setting mode.
2.
Via MODE, go through the settings until bP appears.
3.
Select ON or OFF via CAL
4.
Press MODE to confirm selection
5.
Via MODE, go through the settings until all segments light up
briefly.
6.
The instrument then switches to the measuring mode.