The following functions can be set:
5.3 Opening setting mode
To access setting mode, the instrument must be switched off.
1.
Open setting mode: Hold down MODE + tap ON / HOLD.
5.3.1 Setting the temperature unit (°C or °F)
1.
Open setting mode.
2.
°C or °F appears.
3.
Select temperature unit via CAL.
4.
Press MODE to confirm selection
5.
Via MODE, go through the settings until all segments light up
briefly.
6.
The instrument then switches to the measuring mode.
5.3.2 Switching Auto-Hold on or off
1.
Open setting mode.
2.
Via MODE, go through the settings until AUTO HOLD
appears.
3.
Select On or OFF via CAL.
4.
Press MODE to confirm selection.
5.
Via MODE, go through the settings until all segments light up
briefly.
6.
The instrument then switches to the measuring mode.
5.3.3 Displaying the gradient and offset values stored in the
instrument
1.
Open setting mode.
2.
Via MODE, go through the settings until mV / pH (gradient
and offset values) appear.
3.
Via MODE, go through the settings until all segments light up
briefly.
4.
The instrument then switches to the measuring mode.
5. Operation
27