TESTO 230 Instruction Manual

Ph- and temperature measuring instrument
Hide thumbs Also See for 230:

Advertisement

testo 230 · pH- and temperature measuring instrument
Instruction manual
GlobalTestSupply
www.
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

Summary of Contents for TESTO 230

  • Page 1 230 · pH- and temperature measuring instrument Instruction manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 2: Table Of Contents

    1 Contents Contents Contents ....................3 Safety and the environment ..............5 2.1. About this document ................ 5 2.2. Ensure safety ................... 5 2.3. Protecting the environment .............. 6 Specifications ..................7 3.1. Use ....................7 3.2. Technical data ................. 7 Product description .................
  • Page 3 1 Contents Tips and assistance ................23 8.1. Questions and answers ..............23 8.2. Accessories and spare parts ............26 Appendix ....................27 9.1. Electrode/application assignment ..........27 9.2. Buffer table..................30 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 4: Safety And The Environment

    > Only perform maintenance and repair work on this instrument that is described in the documentation. Strictly follow the specified steps. Use only original spare parts from Testo. > Do not store the product together with solvents. Do not use any desiccants.
  • Page 5: Protecting The Environment

    > At the end of its useful life, send the product to the separate collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal. GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 6: Specifications

    3 Specifications Specifications 3.1. The testo 230 serve the purpose of determining pH-values, redox potentials and temperature in aqueous solutions as well as in semi- solid to solid media. 3.2. Technical data Instrument Feature Values Measurement range pH 0...14, -50...+150 °C, ±1999 mV Resolution pH 0.01, 0.1 °C (°F), 1 mV...
  • Page 7 3 Specifications Feature Values Degree of protection IP 54 (only with plugged in original plug or blanking plug) Housing material Warranty Terms of warranty see www.testo.com/warranty Measuring 24 months instrument 12 months Electrode 12 months Sensor Electrodes/temperature sensors Feature Values pH universal plastic Measurement range: 0...14 pH...
  • Page 8: Product Description

    4 Product description Product description 4.1. Overview 4.1.1. Control elements and connections 1 Display see Display page 10 2 Keyboard Function Switching on/off To call up individual menu levels To hold readings or, with manual input function, set readings To select menu options or, with manual input function, set readings GlobalTestSupply www.
  • Page 9: Display

    4 Product description 3 Battery compartment (on rear) 4 BNC connecting socket for pH and redox electrode 5 Mini-DIN connecting socket for NTC temperature sensor 4.1.2. Display 1 Reading (pH/mV, °C for temperature measurements without pH electrode) 2 DIN/NBS buffer (display in alignment menu) 3 Reading (°C for temperature measurements) 4 Auto-Off display: The circle segments in the display go out in a 2.5 minute interval.
  • Page 10: Connection Variants

    Temperature NTC-sensor with With the NTC-sensor the measurement Mini-DIN plug testo 230 is a fully adequate temperature measuring instrument to cover the range from -50 °C ... +150 °C. GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 11 4 Product description Measurement Probes function Display Redox Redox electrode The redox electrode is measurement used to determine the redox potential of the medium. With the NTC-sensor the temperature can additionally be determined. However, the temperature has no influence on the redox measurement.
  • Page 12: First Steps

    5 First steps First steps 5.1. Commissioning 5.1.1. Installing the sensor Before the instrument is switched on the electrode cable must be connected. 1. Plug the BNC-plug of the electrode into the socket and turn clockwise to lock. 2. Plug the external NTC-sensor with Mini-DIN plug or blanking plug into the socket.
  • Page 13: Calibrating The Instrument

    Single-point calibration serves the purpose of quickly adapting the instrument to the electrode. The following buffers can be used for this purpose: • Testo buffer pH 4, pH 7, pH 10 • DIN buffer pH 4.008 ; pH 6.865 ; pH 9.18 Also pay attention to the application information for the buffer used.
  • Page 14 After max. 10 min. the calibration process is aborted, the electrode is most likely defective, in this case consult our service department. Contact data see back of this document or website www.testo.com/service-contact. The display shows 2. This menu option is needed for two- point calibration, see Two-point calibration page 16.
  • Page 15: Two-Point Calibration

    Two-point calibration serves the purpose of accurately adapting the instrument to the electrode. The following buffer sets can be used for calibration: • Testo buffer set pH 4/7, pH 4/7/10 • DIN buffer set pH 4.008/6.685 , pH 6.865/9.18 Also pay attention to the application information for the buffer used.
  • Page 16 After max. 10 min. the calibration process is aborted, the electrode is most likely defective, in this case consult our service department. Contact data see back of this document or website www.testo.com/service-contact. The display shows 9. Rinse the electrode in water, dab off and place it into the second buffer solution 10.
  • Page 17: Using The Product

    6 Using the product Using the product 6.1. Performing settings testo 230 has a configuration menu in which the following successive settings can be made: 1 Change measuring unit from °C to °F and vice-versa. 2 Switch Auto-Off function on or off.
  • Page 18: Measuring

    6 Using the product Performing a reset (Clear mode) 7. Press SELECT to select yes. > If is selected, the previous values will be maintained. 8. Press HOLD to acknowledge. The previously valid values will be deleted and the ideal values (-58 mv/pH and 0 mV) will be saved in the instrument.
  • Page 19: Temperature Measurement

    6 Using the product 6.2.2. Temperature measurement ✓ NTC sensor connected. 1. Place the NTC sensor into the measuring solution. 2. Press I/O. Press three times. The temperature is displayed. If the display reading is stable, the measuring value is correct. Back to the pH measurement menu: >...
  • Page 20: Auto-Hold Function

    6 Using the product 6.2.5. Auto-Hold function The Auto-Hold function monitors the setting time during a measurement. Once the reading is stable, this status is indicated both visually and audibly. 1. Start measuring process (see Measuring Seite 19). 2. Press to select Auto-Hold.
  • Page 21: Maintaining The Product

    > The instrument is suitable for contact with weak acids and alkaline solutions. After use in aggressive environments the instrument must be cleaned with a damp cloth. For care and maintenance of the pH-electrodes you should read the operating instructions "testo pH-electrode". GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 22: Tips And Assistance

    8 Tips and assistance Tips and assistance 8.1. Questions and answers Question Possible causes / solution Err 1 The pH reading is not in the range 0...14 pH. • No pH-electrode connected. > Connect a pH-electrode. • Incorrectly calibrated. > Recalibrate. •...
  • Page 23 > Rinse the electrodes and repeat the process. If the error occurs again, use a new electrode. Err 7 • Internal instrument fault, measuring instrument defective. > Consult the Testo Customer Service or your dealer. GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 24 24 hours in a solution of thiocarbamide in 0.1M HCI and finally rinse with distilled water. If we have not answered your question, please contact your local dealer or Testo´s Customer Service. Contact data see back of this document or website www.testo.com/service-contact. GlobalTestSupply www.
  • Page 25: Accessories And Spare Parts

    8 Tips and assistance 8.2. Accessories and spare parts Description Article no. Electrodes and sensors pH universal plastic electrode without 0650 2063 temperature sensor, including storage/wetting pH universal plastic electrode with 0650 2064 temperature sensor, including storage/wetting pH glass electrode with temperature sensor, 0650 1623 including storage/wetting cap pH food electrode without temperature...
  • Page 26: Appendix

    9 Appendix Appendix 9.1. Electrode/application assignment Key: - unsuitable, 0 suitable with reservations, + well suitable New type 0650 0650 0650 0650 – – – – designation 2063 1623 0245 2064 Old type Type – Type – Type Type Type Type designation Application...
  • Page 27 9 Appendix New type 0650 0650 0650 0650 – – – – designation 2063 1623 0245 2064 Old type Type – Type – Type Type Type Type designation Application Meat in the puncture Fruit, vegetable in the puncture Hydrofluoric acid containing media Plating-shop waste...
  • Page 28 9 Appendix New type 0650 0650 0650 0650 – – – – designation 2063 1623 0245 2064 Old type Type – Type – Type Type Type Type designation Application Partly aqueous solutions, > 10 % Partly aqueous solutions, < 10 % pH-measurement at temperatures up to 100 °C...
  • Page 29: Buffer Table

    9 Appendix 9.2. Buffer table The table below lists the temperature related changes to the buffers. (DIN buffer acc. to DIN 19266) In the calibration this is automatically accounted for. DIN buffer Testo buffer °C 1.685 4.006 6.865 9.180 4.00 7.00...