Disinfecting The Probe; Changing The Module; Changing Battery - TESTO 206 Instruction Manual

Hide thumbs Also See for 206:
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 20

6.4 Disinfecting the probe

The measuring instrument should not be exposed to disinfectant
for an excessive period of time. Using the TopSafe protective
case can reduce the risk of incompatibility.
1.
Wet a paper towel with disinfectant.
2.
Gently dab the probe with the disinfectant, avoid rubbing.
3.
Ideally, regenerate the probe in electrolyte gel for 12 hours
before carrying out another measurement. (Using alcoholic
disinfectants removes water from the glass membrane)
4.
After each disinfection, carry out calibration with the buffer
solution. (see 5.5 Calibrating the instrument)

6.5 Changing the module

The instrument must be newly calibrated if a module is
changed (See 5.5 Calibrating instrument)!
Instrument must be switched off. Plug contacts in the instrument
should not be touched!
1.
Unscrew the screws at the back of the instrument.
2.
Remove module and attach new module.
There must be rubber sealing rings available for the screws.
Check that the sealing rings are in their correct position.
f
3.
Tighten screws.

6.6 Changing battery

1.
Open battery compartment at the back of the instrument.
2.
Remove spent button cell and insert new button cell (Type
CR2032, 3V). The (+) symbol should be visible.
3.
Close the battery compartment.
6. Service and Maintenance
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

206-ph1206-ph2206-ph3

Table of Contents