Inspection, disinfection, and cleaning
• Do not spray the detector directly with disinfectants or detergents.
Prohibited
• Perform regular inspections.
For safe use of this product and each piece of equipment, perform daily
inspections. In addition, carry out regular inspections and clean the product
at least once a year.
• Always keep this product and other equipment clean and remove all
dust and dirt.
Dust and dirt may cause malfunctions of the equipment included in the
radiography system, such as this product and computers.
• When cleaning the battery pack, turn off the detector and remove the
battery pack from the detector.
• When cleaning the battery pack, wipe it with a cloth slightly damped
with water or diluted neutral detergent.
The battery pack is not protected against liquids. When cleaning the battery
pack, wipe it carefully so as not to spill the detergents onto the electrodes.
• Dry the battery pack completely after cleaning, and attach it to the
detector.
• When cleaning any other parts than the sensor side of the detector,
wipe them carefully so as not to spill the detergents onto the battery
pack attachment portion (electrodes).
• Use water or diluted neutral detergent to clean the surface of the
detector and other equipment. Do not use absolute alcohol, paint
thinner, benzine, or chlorinated solvents. Doing so may damage the
surface of the detector or other equipment.
• Dry the detector completely after disinfecting or cleaning it.
CAUTION
1
Safety information
15