Installation Restrictions; Installation-Related Precautions - Canon AX-C4343W Technical Document

Digital x-ray imager 2.4 ghz/5 ghz
Table of Contents

Advertisement

5) A location where the power source providing the correct power line frequency,
voltage and power used by the product can be supplied
6) A location where a grounding wire and the 3-pin grounding plug of the power cable
can be connected
(3)
After installing the product, ensure that the product packaging materials and other
waste materials generated at the time of installation are disposed of in an
environmentally sound manner.

7.3 Installation restrictions

(1)
Take steps to ensure that the system is not installed near electronic devices or
equipment since noise artifacts occur more readily on the images in environments
where strong electromagnetic fields are present.
(2)
The system must not be installed near devices or equipment which may interfere
with wireless communication.
(3)
When using a wired connection for the product, install the detector and Multi box
with a distance of at least 150 mm between them.
(4)
For use, the cables must be routed without bending, pinching or stretching them or
subjecting them to unreasonable loads. Remodeling of the cables is prohibited.
(5)
In the case of a network connection, the appropriate network address must be
acquired from the network administrator in the installation location, and it must be
set correctly.
(6)
There are no usage restrictions on how the detector is to be positioned. Avoid using
the detector where it will be subjected to extreme bending loads or where it will
easily be dropped.
(7)
Note that the radio frequency channel configured for indoor use may not be usable
in outdoor areas in the 5 GHz band. For further details, refer to "5.1 Wireless
communication specifications" (➞ page 78) in this chapter.

7.4 Installation-related precautions

(1)
Bear in mind that there will be a risk of malfunctioning and other trouble if this
product is used together with any equipment which may generates power supply
noise, supply voltage fluctuations, electromagnetic waves and other disturbances.
(2)
Before proceeding with the installation work, ensure that the static charged in the
bodies of the installation personnel has been released.
(3)
Before connecting or disconnecting the cables, be absolutely sure to turn off the
power of the units in order to ensure safety. In addition, do not touch the power
outlets or connectors with wet hands.
(4)
Based on the wireless channel setting guidelines in "5.1.5 Channel setting guidelines"
(➞ page 79) in this chapter, check out the wireless environment.