Main Specifications - Canon AX-C4343W Technical Document

Digital x-ray imager 2.4 ghz/5 ghz
Table of Contents

Advertisement

8
Specifications

Main specifications

AX-C4343W Detector
Pixel pitch:
Scintillator:
Effective imaging area:
Gray scale:
Attenuation equivalent
of the detector front panel:
Load capacity
Uniform load
(Over the whole area of the
detector surface)
Uniform load
(Effective imaging area)
Local load
(On an area 40 mm in
diameter)
Environmental requirements:
Operation
Storage (unpacked)
Transportation and storage (in packages at point of purchase)*
Applicable grid:
Wireless LAN:
Rated power supply:
Wireless:
Wired:
Dimensions and mass:
54
Temperature:
Humidity:
Atmospheric pressure:
Temperature:
Humidity:
Atmospheric pressure:
Temperature:
Humidity:
Atmospheric pressure:
* Environmental requirements for transportation and storage in packages at point
of purchase are the same for the detector and all optional parts (➞ page 64).
125 μm
CsI (Cesium iodide)
415 x 426 mm
65536 gradations (A/D: 16 bit)
Max. 0.37 mmAl
310 kg or less
150 kg or less
100 kg or less
5°C to 35°C
30% to 80% RH (without condensation)
613 to 1060 hPa
5°C to 40°C
30% to 85% RH (without condensation)
613 to 1060 hPa
–30°C to 50°C
10% to 95% RH (without condensation)
613 to 1060 hPa
34, 40**, 52**, 60** lp/cm (** recommended)
(34 and 40 lp/cm have restrictions. For details, see
page 39.)
Compliant with IEEE 802.11a/b/g/n
9 to 12 V DC, 2.00 A
Powered by the battery pack
Powered by Multi Box
Approx. 460 x 460 x 15.7 mm
Approx. 3.7 kg (incl. battery pack)