ERROR
MESSAGE
BEFORE POWER ON, PLEASE CHECK THE SPECIFICATION AND CONNECTION AGAIN.
SELF-DIAGNOSIS AND ERROR CODE:
DISPLAY
ovfl
Display is positive-overflow (Signal is over display range)
-ovfl
Display is negative-overflow (Signal is under display range)
ADC is positive-overflow (Signal is higher than input
ovfl
high
ADC is negative-overflow (Signal is lower than input
-ovfl
-20%)
eep
fail
EEPROM occurs error
aiCng
pv
Calibrating Input Signal do not process
aic
fail
Calibrating Input Signal error
aoCng
pv
Calibrating Output Signal do not process
aoc
fail
Calibrating Output Signal error
FRONT PANEL:
Relay status
Indication
R L 1
RL 2
2
RL3
RL4
Display screen
?
?
ENT/FN
Operation Key
Numeric Screens
0.8"(20.0mm) red high-brightness LED for 4 2/3 digital present value.
I/O Status Indication
●
Relay Energized: 4 square red LED
RL1
display when Relay 1 energized;
RL2
display when Relay 2 energized;
RL3
display when Relay 3 energized;
RL4
display when Relay 4 energized;
●
External Control Input Energized: 3 square green LED
ECI1
display when E.C.I. 1 close(dry contact)
ECI2
display when E.C.I. 2 close(dry contact)
ECI3
display when E.C.I. 3 close(dry contact)
●
RS485 Communication: 1 square orange LED
COM
will flash when the meter is receive or send data, and
quickly means the data transient quicker.
●
Max/Mini Hold indication: 2 square orange LEDs
M.H
displayed: When the display function has been selected in
CO
Maximum or Minimum Hold function.
Stickers:
Each meter has a sticker what are functions and engineer label
enclosure.
●
H H
Relay energized mode:
●
E.C.I. functions mode:
PV.H(PV Hold) /
Tare /
M.RS(Maximum or Minimum Reset) /
R.RS(Reset for Relay Latch)
●
Engineer Label: over 80 types.
C2-09
-6/8
DESCRIPTION
20%)
Max. Hold status
Comm. status
M.H
C OM
CI11
Control Input
CI2
CI3
status
M.H
Mini. Hold status
°C
?
Engineer Unit
COM
L L
H i
L o
D O
DI(Digital Input)
CS2-T-2010-09-27
(Please check the input signal)
(Please check the input signal)
range
(Please check the input signal)
range low
(Please check the input signal)
(Please send back to manufactory for repaired)
(Please process Calibrating Input Signal)
(Please check Calibrating Input Signal)
(Please process Calibrating Output Signal)
(Please check Calibrating Output Signal)
Operating Key: 4 keys for
Up key /
Down key
Up key
Down key
Shift key
Enter/Fun
key
Pass Word Pcode: Settable range: 0000~9999;
User has to key in the right pass word so that get into【Programming
Level】. Otherwise, the meter will go back to measuring page. If user
forgets the password, please contact with the service window.
Function Lock: There are 4 levels programmable.
●
None
none: no lock all.
●
User Level
user: User Level lock. User can get into
User Level for checking but setting.
●
Programming Level
flash
User can get into programming level for checking but setting.
●
ALL
all: All lock. User can get into all level for checking but
setting.
Front Key Function
●
The
Key can be set to be the same function as the setting of ECI1.
Ex. The ECI1 set to be pVhld and the function [E1=up] set to be
yes in [ eci group]. When user presses
as like as ECI1 close.
●
The
Key can be set to be the same function as the setting of ECI2.
Ex. The ECI2 set to be reLpv and the function [E2=dn] set to be
yes in [ eci group]. When user presses
relative value as like as ECI2 close.
If the front key function has been set, the terminal input for ECI
will be disabling.
REMARK
Enter(Function) /
Setting Status
Function Index
Increase number
Go back to previous
function index
Decrease number
Go to next
function index
Shift the setting
Go back to this function
position
index, and abort the setting
Setting Confirmed
From the function index to
and save to EEProm
get into setting status
eng: Programming level lock.
Key, the PV will hold
Key, the PV will show
Shift(Escape) /
Need help?
Do you have a question about the CS2-T and is the answer not in the manual?
Questions and answers