Calibration; The Calibration Procedure For Et-7015 - ICP DAS USA ET-7000 Series User Manual

Ethernet remote i/o module
Hide thumbs Also See for ET-7000 Series:
Table of Contents

Advertisement

Chapter 1 Introduction


1.8. Calibration

It is not recommended that calibration be performed until
the process is fully understood.

1.8.1. The calibration procedure for ET-7015:

For the ET-7015, each channel should be calibrated separately and only the channel being
calibrated should be enabled during calibration.
1. Warm up the module for at least 30 minutes.
2. Enable the channel you want to calibrate and disable other channels.
3. Set the type code to the type you wish to calibrate.
4. Enable calibration
5. Connect the zero calibration resistor to the enabled channel.
6. Send zero calibration command.
7. Connect the span calibration resistor to the enabled channel
8. Send the span calibration command.
9. Disable calibration
Table 1.8.1 Modbus address for calibration:
Register Points
00830
1
00831
1
00832
1
00833
1
00834
1
00835
1
00836
1
00837
1
00838
1
00839
1
00840
1
00841
1
00842
1
00843
1
00844
1
Enable/Disable Calibration
0=Disabled
1=Enabled
Zero calibration command of ch0, 1 = run.
Span calibration command of ch0, 1 = run
Zero calibration command of ch1, 1 = run.
Span calibration command of ch1, 1 = run
Zero calibration command of ch2, 1 = run.
Span calibration command of ch2, 1 = run
Zero calibration command of ch3, 1 = run.
Span calibration command of ch3, 1 = run
Zero calibration command of ch4, 1 = run.
Span calibration command of ch4, 1 = run
Zero calibration command of ch5, 1 = run.
Span calibration command of ch5, 1 = run
Zero calibration command of ch6, 1 = run.
Span calibration command of ch6, 1 = run
ET-7000 AIO User Manual, Apr. 2009, V1.04,
Description
EMH-011-104----------40
Access Type
R/W
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)
W (Pulse)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Et-7015Et-7017Et-7017-10Et-7018z

Table of Contents