CP 20_________________________________________________________________Section 4
Please note that the default Real-Time ECG View (as pictured below) is not available in the Acquired ECG
View for navigation purposes.
Real-Time ECG View
Note: new softkey labels are available in the Acquired ECG View.
Note: functions are not available during acquisition.

Printing

If the auto-print configuration option is enabled, an ECG is printed following acquisition. Regardless of
the auto-print configuration setting, a manual printout is possible from the Acquired ECG View by
selecting Copy (F4).
If the auto-print configuration option is disabled, the 10-second preview assists in ensuring a quality ECG
acquisition prior to printing. When you acquire an ECG, the cardiograph captures the previous 10 seconds.
The relationship between the display and the printout is the same – what is displayed in the ECG
Acquisition View is what will be printed.
To manipulate the print format of the acquired ECG, regardless of the plot format configuration setting,
select Fmt (F2). The following function keys become available:
Select the function key corresponding to the desired print format. The Acquired ECG View is then
displayed and to make an ECG printout copy in the new plot format; select Copy (F4). Select Exit (F6)
to return to the Real-Time ECG View.
Acquired ECG View
4-6