Configuration Page 5 (for Optional Communication Module)

Default XMT Media
This will define which transmission mode will be used by the user. IF RS 232 is selected, the baud rate selected in
the configuration will be used. If modem is selected, the system will use the phone number stored in the page 1 of
the configuration screen If WLAN is selected, pages 5 and 6 are used to set the parameters.
DHCP (Dynamic Host Communication protocol)
Defines whether DHCP will be used to obtain an IP address. If DCHP is Yes, there is no need to enter the IP address
(the network will automatically and dynamically assign an IP address. If DHCP is No, the user must enter the IP
address, Def Gateway and Sub Net Mask.
NOTE: all parameters related to network connection must be entered based on inputs provided by the IT
manager of the facility where the unit will be used.
IP Address
Define the fixed IP address used by the CP 20 for network transmission. (If DHCP is not selected)
Def Gateway
Enter the address of the default gateway. (If DHCP is not selected)